Operating the Sliding Sash 
The counterbalanced, anti-racking sash mechanism requires only a few 
pounds of force to move the sash up or down. You can open or close the sash 
smoothly with one or two hands positioned on either handle. 
The sash position alarm and safety interlock system senses the sash position 
and acts appropriately. The procedure station has been programmed to 
operate at either an 8- or 10-inch (203-254 mm) sash opening, depending on 
model. Raising the sash above its operating height will activate the audible 
and visual alarms. The audible alarm can be temporarily muted (for 
approximately five minutes) by depressing and releasing the OK/Mute 
button. Closing the sash back to its operating position will reset the alarm and 
defeat the muting of the alarm. The safety interlock system senses when the 
sash is closed and allows the optional ultraviolet (UV) lamp to operate only 
when the sash is closed, to protect the operator from irradiation. 
Starting the Procedure Station 
1.  To start the cabinet, raise the sash until its bottom edge aligns with the 
sash position label on the left corner post. The decal is shown in 
Figure 5-5. 
2.  Press the blower button to start the unit. The unit will display a 
standby screen for approximately 90 seconds to allow the cabinet to 
reach proper operating conditions. If the alarm sounds, recheck the 
sash position. If the sash is too high, the sash audible alarm and the 
LCD display will indicate the sash is too high. 
3.  To turn the UV light on, the sash must be completely closed to 
prevent the escape of any UV radiation. Push the UV light button to 
activate the UV light. 
Note: The sash must be completely closed for the UV light to activate. 
Sash position label
Figure 5-5 