User guide
Chapter 5: Using The Procedure Station 
Product Service 1-800-522-7658 
30 
Original instructions 
Configuring the Procedure Station 
In the first MyLogic screen, use the ▲ and ▼ 
buttons on the touchpad to highlight configure my 
Logic for use -it will turn white when selected. Press 
OK to enter the first configuration screen: 
The first screen gives you the option of activating 
the Smart-Start option for the blower; if you want 
the cabinet blower to start every time you raise the 
sash, select start, and then press OK. If not start is 
selected, then the blower must be manually started 
from the keypad. When OK is pressed, the next 
configuration screen will appear. 
The next screen gives you the option of activating 
the Smart-Start option for the fluorescent light; if 
you want the lights to turn on every time you raise 
the sash, select turn on, and then press OK. If stay 
off is selected, then the lights must be manually 
turned on from the keypad. When OK is pressed, the 
next configuration screen will appear. 
If you want the cabinet blower to run slowly, 
maintaining reduced airflows every time you close 
the sash, select go into Night-Smart mode and then 
press OK. If stop is selected, then the blower will 
stop when the sash is closed. When OK is pressed, 
the next configuration screen will appear. 
If your cabinet is configured for a UV light, then 
you will see the next two screens; if you want the 
UV lamp to turn on every time you close the sash, 
select go into Night-Smart mode, and then press OK. 
If stay off is selected, then the UV light will not turn 
on when the sash is closed. When OK is pressed, the 
final configuration screen will appear. 