User guide
Chapter 5: Using The Procedure Station 
Product Service 1-800-522-7658 
34 
Original instructions 
Change UV Lamp Life 
In this screen, you can set the number of operating hours before receiving the 
replace UV lamp warning. For most UV lamps, the output of UV light 
decreases at a constant rate. Typically, after 6,000 hours of operation the 
lamp will output 80% of the UV light it did when it was new. This option 
allows you to set operational life of the UV lamp, in 100 hour increments. 
From the UV Settings menu screen, use the ▲ and ▼ 
buttons on the touchpad to highlight the Set UV Life 
option-it will turn white when selected. Press OK to 
enter the Set UV life screen. Use the ▲ and ▼ buttons 
to increase or decrease the lamp life in 100 hour 
increments, until the desired value is displayed, and then 
press OK. The screen will then return to the UV 
Settings screen. 
The Service Menu Screens 
Note: The Service Menu screens are reserved for use by trained certification 
personnel as part of the certification or service procedures. All of the screens 
have a yellow background, and those portions of the Service Menu screens 
that can alter the performance of the cabinet are password protected. If you 
have any questions about these screens, contact Labconco’s Product Service 
Department at 1-800-821-5525 or www.labconco.com for assistance. 