Timer Operation 
NOTE: The timer button allows activation of an interval (countdown) or 
stopwatch (elapsed) timer. The timers cannot be operated 
simultaneously. 
To access the main timer menu, press the Timer button anytime during 
normal operation. The main timer menu is shown on the LCD display. Use 
the ▲ and ▼ buttons to highlight the Interval or Stopwatch Timer. Press the 
OK button to select the highlighted timer function. 
Interval Timer Operation 
1.  When selected, the Interval Timer menu is displayed on the LCD. 
The timer defaults to 0:00:00 (hours:minutes:seconds). 
2.  Press and hold the ▲ or ▼ buttons to increase or decrease the timer 
interval. 
3.  When the proper interval is entered on the display, press the OK 
button to start the timer. 
4.  When the timer reaches 0:00:00, an audible alarm will sound, and the 
timer will reset itself and repeat the countdown. 
5.  Press the OK button to pause the timer. 
6.  Press the Menu button to clear the interval timer and return to the 
main timer menu. 
Stopwatch Timer Operation 
1.  When selected, the Stopwatch Timer menu is displayed on the LCD. 
The timer defaults to 0:00:00. 
2.  Press the OK button to start the timer. 
3.  Press the OK button again to zero the timer. 
4.  Press the Menu button to clear the stopwatch timer and return to the 
main timer menu. 
If An Airflow Alert Activates 
The most common causes of an Airflow Alert are: 
•  Blockage of the inlet grilles or exhaust outlet. 
•  Removal of the work surface or grille during operation. 
Resetting the Airflow Alert System 
The Airflow Alert automatically resets to normal operation once the motor 
speed has stabilized. 