User guide
Chapter 5: Using The Procedure Station 
Product Service 1-800-522-7658 
36 
Original instructions 
Working In the Procedure Station 
Note: A more thorough review of using a Procedure station can be found in: 
Biosafety in Microbiological and Biomedical Laboratories 
(BMBL), Published by the Centers for Disease Control and Prevention 
(www.cdc.gov/biosafety/publications
). 
Planning 
•  Thoroughly understand procedures and equipment required before 
beginning work. 
•  Arrange for minimal disruptions, such as room traffic or entry into the 
room while the cabinet is in use. 
Start-up 
•  Turn off UV light if included. 
•  Slowly raise the sash until the bottom of the sash aligns with the sash 
indicator decal located on the left side of the work area. 
•  Turn on the fluorescent light and cabinet blower if the Smart-Start
features have not been activated. 
•  Check the air grilles for obstructions. 
•  Allow the cabinet to operate until the display screen is shown. 
Warming up Screen  Display Screen 
•  Wash hands and arms thoroughly with germicidal soap. 
•  Wear appropriate personnel protective equipment (PPE). 
Wipe-Down
•  Raise the sash to its full open position (approximately 21.75 inches or 
552 mm). Mute the alarm by depressing the "OK/Mute" switch. 
•  Wipe down the interior surfaces of the cabinet with 70% ethanol, or a 
suitable disinfectant, and allow to dry. 