The common service operations necessary to maintain the Procedure Station 
for peak performance are listed below. 
Note: This manual covers operation and maintenance operations for the 
owners/users of the PuriCare Procedure Stations. Complete certification 
procedures, service operations and specifications are published in a separate 
publication Technical Manual: Purifier
®
 Logic
®
+ Biosafety Cabinets. This 
manual is available from Labconco’s website: www.labconco.com. A 
complete certifier service kit is available to qualified certifiers from 
Labconco. Call Labconco’s Product Service Department at 800-821-5525 or 
816-333-8811. 
  Do NOT contact blower wheel while still in motion. 
NE PAS être en contact avec la roué du ventilateur tant qu’il est 
en marche. 
Routine Maintenance Schedule 
Weekly 
•  Wipe down the interior surfaces of the cabinet with a suitable 
disinfectant, or 70% ethanol, and allow to dry. 
•  Using an appropriate glass cleaner, or LabSolutions
TM
 Glass & 
Surface Wipes, Labconco part # 1570000; clean the sash and the 
surface of the UV lamp, if so equipped. 
•  Operate the cabinet blower, noting the percent filter life remaining in 
an operational log. 
Monthly (or more often as required)
•  Using a damp cloth, or LabSolutions Glass & Surface Wipes, 
Labconco part # 1570000; clean the exterior surfaces of the cabinet, 