User guide
Chapter 6: Maintaining The Procedure Station 
Product Service 1-800-522-7658 
40 
Original instructions 
particularly the front and top of the cabinet, to remove any 
accumulated dust. 
•  Disinfect and lift the work surface. Surface disinfect the area beneath 
the work surface with a suitable disinfectant, or 70% ethanol, and 
allow to dry. Check the towel catch and prefilter for retained 
materials. Replace prefilter if necessary. 
•  Check all service valves, if so equipped, for proper operation. 
•  Check the UV and fluorescent light hourmeters, and record their 
readings in an operational log. 
•  All weekly activities. 
Semiannually or Annually 
•  Have the cabinet re-certified by a qualified certification technician. 
•  All monthly activities. 
Service Operations 
Work Surface Removal: 
Note: The work surface must be thoroughly decontaminated before 
removing it from the cabinet. 
1.  Lift the front edge of the work surface straight up by grasping the knob 
handles at either front corner. 
2.  Pull the work surface straight out, letting its rear edge rest on the center 
support underneath. 
3.  Reinstall the work surface by resting the bottom on the center rail while 
pushing it back into the cabinet. Be sure to engage the tabs on the back 
corners of the work surface with the slots on the rear wall of the work 
area. 
Front Grille Removal:
Note: The grille must be thoroughly decontaminated before removing it. 
1.  Remove the work surface as described earlier. 
2.  At one end of the grille, grip the front of grille with one hand, and the 
back with the other hand. Pivot that end of the grille upward and 
inward, paralleling the angle of the sash, as shown in Figure 6-1. 
3.  Pull the other end of the grille up and away from the bottom edge of 
the cabinet. 