Chapter 6: Maintaining The Procedure Station 
Product Service 1-800-522-7658 
44 
Original instructions 
Changing the Fluorescent Lamps: 
1.  Unplug the cabinet or turn off the System Reset Switch located on the top 
of the cabinet. 
2.  Remove the front dress panel as noted in Figure 6-3. 
3.  Remove the fluorescent lamps by pulling the lamp sockets straight off 
each end of the lamp, and releasing both lamps from the spring clips that 
secure them in place. 
4.  Install the new lamps by reversing the removal procedure. 
Changing the Optional UV Lamp: 
Note: For optimum performance, the UV lamp should be changed on an 
annual basis, or as indicated by the UV lamp timer. 
The UV lamp and the work area of the cabinet must be thoroughly 
decontaminated before removing the lamp. 
1.  Start the cabinet and let it operate for 5 minutes. 
2.  Raise the sash to its full open position. 
3.  Thoroughly surface decontaminate the UV lamp and the work area of the 
cabinet. 
4.  Unplug the cabinet or turn off the System Reset Switch, located on the 
top of the cabinet. 
5.  Remove the UV lamp by rotating it 90 degrees and lifting it straight up 
and out of its sockets. 
6.  Install new lamp by reversing the removal procedure. 