User guide
Chapter 7: Troubleshooting 
Product Service 1-800-522-7658 
48 
Original instructions 
PROBLEM CAUSE  CORRECTIVE ACTION 
Airflow Alert goes 
off and/or there is a 
slight decrease in 
filter life remaining 
gauge 
HEPA filter loading  The gauge reading steadily decreases as 
the cabinet is used. 
  Blockage of the 
return air slots or 
grille 
Check all return air slots and grilles to 
ensure that they are not blocked or 
restricted. 
  Blockage of the 
exhaust outlet 
Ensure that the exhaust outlet is not 
blocked or restricted. 
 Blockage or 
restriction under the 
work surface 
Ensure that the towel catch and plenum 
beneath the work surface are 
unobstructed. 
Contamination of 
work in the cabinet 
Improper technique 
or procedure for the 
procedure station 
See “Use of the cabinet” section in the 
manual. 
  Restriction of the 
return air slots or 
grille – blockage of 
the exhaust outlet 
Ensure that all return air slots, grilles 
and the exhaust outlet are unobstructed. 
  External factors are 
disrupting the cabinet 
airflow patterns or 
acting as a source of 
contamination 
See “Working in the Procedure station” 
section of this manual. 
  Cabinet is out of 
adjustment/HEPA 
filter(s) are defective 
Have cabinet recertified. 