The common service operations necessary to maintain the Procedure Station
for peak performance are listed below.
Routine Maintenance Schedule
Weekly
Using 70% ethanol, or a suitable disinfectant, surface disinfect the
inside of the cabinet and the work surface.
Using an appropriate glass cleaner, or LabSolutions Glass & Surface
Wipes, Labconco part # 1570000; clean the sash and the surface of
the UV lamp, if so equipped.
Operate the cabinet blower, noting the percent filter life remaining in
an operational log.
Monthly (or more often as required)
Using a damp cloth, or LabSolutions Glass & Surface Wipes,
Labconco part # 1570000; clean the exterior surfaces of the cabinet,
particularly the front and top of the cabinet, to remove any
accumulated dust.
Disinfect and lift the work surface. Surface disinfect the lower
plenum with a solution of 70% ethanol, or a suitable disinfectant.
Check the towel catch and prefilter for retained materials. Replace
prefilter if necessary.
Check all service valves, if so equipped, for proper operation.
Check the UV and fluorescent light hourmeters, and record their
readings in an operational log.
All weekly activities.
Semiannually or Annually
Have the cabinet re-certified by a qualified certification technician.
All monthly activities.