User guide
General
The LABEC Automatic Autoclave is designed and constructed to meet the requirements of the
S.A.A. boiler Code and has been approved by the Boiler Inspectors Branch of the N.S.W.
Department of Industrial Relations ‘Workcover Authority’, to operate at pressures up to 138 K.P.A.
Regulations require that on installation the autoclave be examined by a licensed boiler inspector
andthiscertificateofinspectionbedisplayedunderglassneartheautoclaveandrenewedannually.
ApprovalNumbers
NSW
QLD
SA
WA
TAS
AA9
6766P87
WA94
AA13/AA16
674P90
BV18716
PV86636/7
WA9489
TA4946
AA18/AA20
673P90
BV18717
PV86636/7
WA9490
TA4947
Heating
Heating of the autoclave is by means of stainless tubular element located in the base of the
autoclavevessel.
DescriptionofControls
SeeAppendixAAttached
TemperatureControl
Labec autoclaves are fitted with solid state proportional action temperature controllers which operate
from a sensitive thermocouple or Rtd inserted in the working space of the chamber. The controller
has been calibrated at 121°C and before any adjustments to temperature settings are made allow
the autoclave temperature to stabilise for at least one half hour. For the digital controller fitted
please read the enclosed operating instructions when setting the controller temperature. Note: if
you intend using the autoclave at temperatures other than the preset 121°C, you need to
adjust the ‘AL4’ value, using the up or down arrow, set this value equal to the new
temperature at which you use the autoclave (you must do this in LVL2 in the controller).
Now change the temperature using up and down arrows to the new temperature setting
andpressstart.(AL4isthevaluewhichwhenreachedwillactivatethetimer).
ImportantSafetyInstructions
Caution
PleaseobservethefollowingsafetymeasuresbeforeusingyourLABECequipment.
1. Ensure pressure inside the autoclave is reading zero KPA on the pressure gauge before
openingthelid.
2. These autoclaves are NOT FLAME PROOF and under no circumstances should
inflammable,combustibleorexplosivematerialbeplacedintheautoclave.
3. Low ignition temperature materials and those materials which give off inflammable or
explosivevaporsshouldnotbeplacedintheautoclave.
4. Avoidheatingsubstanceswhichgiveoffcorrosivevapor.
5. Users are advised of the dangers of heating combustible materials. The manufacturer can
recommendspecialtypesofelementswhichwillprohibittheautoclavestemperature
6. Observe those rules pertaining to wiring and installation of electrical appliances as
recommendedbythelocalsupplyauthority.
LaboratoryEquipmentPtyLtdInstructionManualAutoclaveAA9BenchTopModel