User guide
Maintenance
The cabinet is finished with stainless steel and to maintain appearance should be wiped over with a
non abrasive cleanser. The chamber is manufactured from stainless steel and may be cleaned with
a solvent and if grained stainless steel chamber a stainless steel scratch pad may be used. The
padshouldberubbedinthesamedirectionasthegrainpatternofthestainlesssteel.
The silicone gasket should be cleaned with detergent only, ensuring it is dried completely after
washing. To prolong life it may be stored at cool temperatures in a container of water. Gasket
lubricantisavailabletoprovideprolongedlifeandbettersealingonthegasket.
The locking pins should be regularly greased with high temperature grease to reduce binding in the
guides.
WARNING:THISAUTOCLAVESHOULDNEVERBECONNECTEDTOMAINSWATERSUPPLY
Troubleshooting
SYMPTOM
REMEDY
NoPower
(IndicatorLightisoff)
1. Check autoclave is plugged in and power
switchedon.
2. Ensure mains power supply point is
functioning by using a test appliance on
powersocket.
3. CheckinternalRCDhasnottripped.
Failuretoheatormaintaintemperature
(Indicatorlightison)
1. Thetemperaturecontrollerison.
2. The water level is above the low water
sensor and the low water indicating lamp
isoff.
Thetimerdoesnotstart
1. Check the AL4 value is equal to set value
asthisvalueactivatesthetimer.
If the fault cannot be found call your distributor or the manufacturer quoting the serial number of the
autoclavefromthemanufacturerslabel.
DeclarationofConformity
Each product is thoroughly inspected and tested to not only ensure that it meets the specifications
provided, but to also meet Australian Electrical Standard AS3820 and EMC Standard
AS/NZ1044:1995,and therefore being accredited with a C Tick label. Labec Autoclaves are
constructed in accordance with Australian Standard 1210 and are approved by Workcover Authority
ofNSWwithreciprocalapprovalbyequivalentbodiesinallstatesofAustralia.
LaboratoryEquipmentPtyLtdInstructionManualAutoclaveAA9BenchTopModel