Manual

An independent temperature checks of temperature in the chamber and the displayed
temperatureontheTemperatureControllershouldbecarriedout.
Allalarmsshouldbetested.
5TroubleShootingandAfterSalesService
Some unexpected failures may arise during running. Please pay attention to the running status
of your ULT freezer. If any problems arise, please check and solve it according to the following
trouble shooting list. If the problem is not solved by any of these, please contact our Service
Departmentwiththeserialnumberoftheunit.
BeforeaServiceVisit,pleasecleananddisinfectthefreezer.
Problem
PossibleCause
ULT freezer does not
operate
Poweroutagehasoccurred.
Theplugisfaultyornotsecurelyplugged-in.
Fuseisblown.
Voltageistoolowortoohigh.
Compressor does not
work
Powerincontrolpanelisoff.
Temperaturesettingiswrong.
Temperature is higher
thansetting
Frostininnerwallistoothick
Thedoorisnotclosedproperlyorisopenedtoofrequently.
Chamberisoverloadedwithproduct.
Ambienttemperatureistoohigh.
LoudNoises
Freezerisnotlevel.
Freezercomingintocontactwithwallorobject.
Belowarewithinnormaloperationparameters:
Therearesomelightclashingnoiseswhenthecompressorstartsupandstops.
During rain and high humidity some frost may accumulate on the surfaces of cabinets. Please
wipeawaywithdryrag.
DO NOT open the door twice within 5 minutes. This may cause the door to be difficult to open
duetonegativepressure,causedbyhotairletinbyopeningdoorcontracting.