Owner's manual
Section 
6
CONTROLS OVERVIEW 
6.1 Power Switch: The main power ON/OFF switch controls all power to the unit. It 
must be on before any systems are operational. The green switch will be lighted 
when ON. 
6.2  Main Temperature Control: This control consists of the digital display and 
Up/Down buttons for inputting set point temperatures and calibration. 
6.3 Safety Thermostat: The Safety is completely independent of the Main Controller. 
The Safety guards against any failure of the Main Controller that would allow 
temperature to rise past set point. If temperature rises to the safety set point, the 
Safety takes control of the heating element and allows continued use of the 
incubator until the problem can be resolved or service can be arranged. 
6.4 Fan Switch: This operates the turbo fan inside the chamber for fast heat up times 
and recovery. The switch is lighted when activated. 
6.5 Safety Light: This pilot light comes ON when the Safety Thermostat is activated. 
 Under normal operating conditions this pilot light should never be on. 
6.6 Circuit Breaker: The breaker is manually resettable and offers protection against 
power source variations. Protection is in addition to the automatic high-temperature 
limit designed into the heating element itself. This is located at the rear of the unit. 
9