User Manual

Adapter for 0.5ml tubes (pk./6)
Order no. C-1205
Tube measurement 8 x 30mm
Max. speed 15,000rpm
Centrifuging radius 7.53cm
RCF (g-value) 18,942 x g
Adapter for 0.4ml tubes (pk./6)
Order no. C-1206
Tube measurement 6 x 47mm
Max. speed 15,000rpm
Centrifuging radius 8.4cm
RCF (g-value) 21,200 x g
Adapter for 0.2ml tubes (pk./6)
Order no. C-1222
Tube measurement 6 x 21mm
Max. speed 15,000rpm
Centrifuging radius 7.03cm
RCF (g-value) 17,684 x g
3.2 Rotor maintenance
The rotor should be cleaned thoroughly after each use. Thorough cleaning must be
performed when spinning samples containing phenol or phenol chloroform.
Periodically inspect the rotor for dents, dings, scratches, discoloration and cracks. If any
damage to the rotor is found, discontinue use of the rotor immediately and replace.
3.3 Removing and installing the angle rotor
The Prism comes complete with a standard 24 place rotor installed. To remove the rotor
for cleaning, remove the rotor securing screw from the motor shaft by turning the screw
counterclockwise, using the rotor wrench. Lift the rotor directly upward in a straight
vertical motion.