Instruction Manual

Error no.: 70Error no.: 70
Error no.: 70Error no.: 70
Error no.: 70
Interface of the frequency converterInterface of the frequency converter
Interface of the frequency converterInterface of the frequency converter
Interface of the frequency converter
Reason: Communication of controller, power board, interface cable
and frequency converter is not working
Action: Call service
Error no.: 82-83Error no.: 82-83
Error no.: 82-83Error no.: 82-83
Error no.: 82-83
Cutoff power board - frequency converterCutoff power board - frequency converter
Cutoff power board - frequency converterCutoff power board - frequency converter
Cutoff power board - frequency converter
Reason: Overcurrent or undervoltage due to power supply fluctuations
Action: Restart the unit, take care that the power supply is stable
Error no.: 84Error no.: 84
Error no.: 84Error no.: 84
Error no.: 84
Overtemperature at the drivingOvertemperature at the driving
Overtemperature at the drivingOvertemperature at the driving
Overtemperature at the driving
Reason: Temperature at the converter or motor too high
Action: Switch off the centrifuge. Wait for about 15 min. and switch
the unit on again
Error no.: 85-87Error no.: 85-87
Error no.: 85-87Error no.: 85-87
Error no.: 85-87
FailuresFailures
FailuresFailures
Failures
Reason: Internal defect
Action: Call service
Error no.: 90Error no.: 90
Error no.: 90Error no.: 90
Error no.: 90
Emergency lid releaseEmergency lid release
Emergency lid releaseEmergency lid release
Emergency lid release
Reason: The centrifuge lid has been opened by the emergency lid release
during the run
Action: Close centrifuge lid. Danger of accident!
Reason: Control switch of the lid lock is defective
Action: Call service
Error no.: 94Error no.: 94
Error no.: 94Error no.: 94
Error no.: 94
Voltage loss during runVoltage loss during run
Voltage loss during runVoltage loss during run
Voltage loss during run
Reason: The power supply is below the tolerance for a short moment
Action: Wait till stillstand of the rotor. Open centrifuge lid after the yellow
LED "lid" is lighting. Switch off and on the main switch.