Owner's Manual
Table Of Contents
- 1 Introduction
- 2 Quick Start
- 2.1 Setting Up the Equipment
- 2.2 Gathering Supplies and Equipment
- 2.3 Preparing the Urocap NXT
- 2.4 Preparing the NXT Go Pump Hub
- 2.5 Preparing the Roam NXT
- 2.6 Preparing the UPP Puller NXT (Optional)
- 2.7 Setting Up Consumables and Preparing the Patient
- 3 How to Perform a Study
- 4 Synergy Software
- 5 Maintenance and Service
- 5.1 Cleaning and Disinfection
- 5.2 Preventive Maintenance
- 5.3 Verifying Calibration
- 5.4 Configuration
- 5.5 Computer Virus Protection
- 5.6 Disposing of Product After Use
- 5.7 Environmental Consideration of Waste Disposal
- 5.8 Backups
- 5.9 Troubleshooting Guide
- 6 Hardware
- 7 Appendix
![](/manual/laborie-medical-technologies/tdoc3/owner-s-manual-english/images/img-103.png)
Hardware
NXT Go Owner’s Manual, LBL-00145, MAN2030 102
Figure 82: Roam NXT and PIM NXT for Fluid Controls
Replacing the PIM NXT
Follow the instructions provided below for replacing the PIM NXT.
IMPORTANT: The PIM NXT is not intended to be interchangeable during regular use.
When the PIM NXT is connected, directly or through use of the Roam NXT to PIM NXT
Connector Cable, to the Roam NXT it should only be detached for cleaning or maintenance.
6.7.2.1 Removing the PIM NXT
To remove the PIM NXT, follow the instructions provided below.
1. Use a flat head screwdriver to loosen the screw(s) connecting the PIM NXT and
Roam NXT without fully detaching.
a. When directly connected, loosen the screw located at the back of PIM
NXT.
b. When using the connector cable, loosen the screw located at the joint
of the connector cable with the PIM NXT and the joint of the connector
cable with the Roam NXT.
2. Gently pull the PIM NXT away from Roam NXT base until it disconnects from
the base (Figure 83).
Figure 83: Roam NXT with a Detached PIM NXT
6.7.2.2 Attaching New PIM NXT
To attach a new or existing PIM NXT, follow the instructions provided below.
Using a flat head screwdriver, loosen the screw(s) used to connect the PIM NXT
and Roam NXT without fully detaching.
Battery LED
Status
Roam NXT
LED Status
EMG Port
PIM NXT for Fluid to Medex
Transducer Cables
Interlocked Roam NXT, PIM NXT for
Fluid, and Cables - Front View
Clip
PIM NXT for Fluid – Front View
Pressure Channel
Connectors
Disposable Cartridge
Medex LogiCal®
Pressure Transducer
Status DRAFT Effective