NXT Go Owner’s Manual, LBL-00145, MAN2030 16
Warnings:
The NXT Urodynamic System contains magnets which can interfere with control of a
pacemaker or ICD (Implanted cardioverter-defibrillator). Follow instructions from the
manufacturer of such devices and ensure that the Roam NXT, Urocap NXT and charger
cable connectors are always kept at a sufficient distance from any implanted device.
Only LABORIE trained technicians may service the unit.
Do not use if the device packaging has been opened, damaged, or if it presents any fault
due to improper transport, storage, or handling that could in any way hamper its use.
The NXT System is intended for install by trained personnel ONLY.
Do not immerse the NXT System or any of its modules in water or any other liquids.
Do not place your fingers or any body parts inside the pump when the pump rollers are
moving.
Do not use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous
oxide.
Do not use electromyography (EMG) simultaneously with high frequency surgical
equipment.
Do not operate the NXT System near shortwave or microwave therapy equipment as
proximity may produce instability in the APPLIED PARTS.
Do not allow applied parts linked to a PATIENT CONNECTION, but not connected to the
PATIENT, to contact other conductive parts including those connected to protective earth.
Preserve the patient connection electrical isolation. Do not place the printer in the patient
environment.
Do not reuse single-use accessories. After use, dispose of single-use accessories in
accordance with local regulations.
Re-use, reprocessing, or resterilization of disposables can lead to device failure and create
a risk of cross-infection and/or cross transmission of infectious disease(s) from one
patient to another.
NXT System Accessories are not intended to be defibrillation-proof.
Improper processing of reusable items prior to initial use or reuse can result in cross
infection and/or cross-transmission of infectious disease from one patient to another.
Risks associated with catheterization increase immediately after urethral or bladder
surgery. The performance of Urodynamic studies with post-operative patients is at the
discretion of the physician.
Medical care professionals performing Urodynamic studies should be prepared to
recognize and treat symptoms associated with ureteral catheterization, vasovagal
syncope (fainting), and Autonomic Dysreflexia.
To avoid potential patient injury never clean or service any module or component of the
NXT System while in use with a patient.
The use of the Medical Electrical equipment is restricted to one patient at a time.
Do not use the NXT System in the presence of a magnetic resonance imaging system as
the NXT System may contain ferromagnetic objects that pose a risk to the patient in the
presence of a magnetic core.