Owner's Manual

Table Of Contents
Quick Start
NXT Go Owner’s Manual, LBL-00145, MAN2030 24
2.7 Setting Up Consumables and Preparing the
Patient
Synergy uses Consumable Traceability to track consumables used during studies. Prior
to beginning consumable setup, ensure to start a study. For instructions on starting the
study refer to the Start Study Button section on page 61.
Complete consumable and patient setup in the preview phase of the study. Catheters and
tubing must be scanned into Consumable Traceability to proceed from the preview phase.
Consumable information will show in the Consumable Traceability tab available in the
Preview window once entered or scanned. For more information on Consumable
Traceability refer to the Procedure Preview Phase section on page 62.
Infusion Pump Tubing Setup
LABORIE Infusion Pump Tubing is intended for single use only. Do not re-use, reprocess, or
resterilize pump tubing. Use of the Infusion Pump Tubing outside of its intended use may
affect pump accuracy. Do not run the pump at a speed exceeding the limitation of the
catheter.
IMPORTANT:
Always use only LABORIE Infusion Pump Tubing to ensure a properly and
accurately functioning system.
Catheters, the saline bag, and the Infusion Pump Tubing are consumables intended
for single use with a single patient only; they must be replaced after every use.
To maintain patient safety during normal use, the maximum pump speed is preset
to 150ml/min.
Always refer to the Instructions for Use provided with the consumable deployed for
information on proper use of the device.
To setup the pump tubing, follow the steps provided below:
Scan the SmartSense tag, located on the packaging, against the reader located on
the NXT Go Pump Hub by holding proximity for 3 seconds. Refer to the NXT Go
Pump Hub section on page 98 for a visual of the SmartSense tag reader.
Hang a bag of saline solution. Refer to the Infusion Volume Transducer section on
page 100. Insert the pump tubing spike into the bag.
Close the tubing clamp to block the tubing and to stop the flow.
Squeeze and release the flexible drip chamber until fluid flows.
Open the tubing clamp and flush the line completely.
Open the pump door and position the compressible portion of the pump tubing
across the rollers; close the pump door.
NOTE: Do not stretch the compressible portion of the pump tubing during insertion as
the pump controls the tautness of the tubing insert.
To remove air bubbles from the pump tubing with Synergy, click the Prime Pump Tube On
button. Click the Prime Pump Tube Off button to turn off the pump. Synergy
automatically displays pump activation markers.
Attach the luer connector of the tubing to the fill connection of a prepared catheter.
Status DRAFT Effective