How to Perform a Study
NXT Go Owner’s Manual, LBL-00145, MAN2030 36
3.3.2.3 Filling Phase
Click the Start Filling Phase button . The Filling phase
window will display.
NOTE: Pause the study by clicking the Pause button and resume the study
by clicking the resume button.
If necessary, the Pabd channel value can be set to equal the Pves channel value
by pressing the Equalize Pdet button.
Click the Pump On button to start the Infusion Pump and begin filling the
bladder. Click Smart Pump Up or Smart Pump Down buttons to increase or
decrease the pump speed, respectively.
NOTE: To change the infusion bag click the Change Infusion Fluid button. A popup
window will appear, and the study will pause. If desired, update the Fluid Density setting
using the dropdown menu available on the popup window and click OK to resume the study.
Fluid density options must be configured in the Workflow Configuration window before
starting the study. Refer to the Customizing Workflows section on page 56.
Encourage the patient to give feedback on sensation or urge. Mark relevant events
(e.g.: First Sensation, First Desire, Strong Desire, etc.) by clicking the First
Sensation drop-down menu and selecting the relevant event. Refer to Figure 23.
Figure 23: Pressure Flow Filling Phase
When the patient starts to feel discomfort or fullness, stop the Infusion Pump by
clicking the Pump Off button in the Synergy software. Click the Start Voiding
Phase button to move to the next phase.