NXT Go Owner’s Manual, LBL-00145, MAN2030 74
Flow Phase
Blue=0 points
Yellow=+5 points
Green=+10 points
Volume Voided
Pediatric
No uroflow
volume recorded
Volume voided > 0
Volume voided ≥
50% EBC
Volume Voided
Adult
No uroflow
volume recorded
Volume voided > 0
Volume voided >
150 ml
PVR
No PVR value
entered
Not Possible
Any PVR value,
including 0 ml
Beaker
Beaker not
empty at start of
flow (Beaker
weight >= 25
gm)
Beaker empty at start
of flow (Beaker
weight < 25 gm)
Not Possible
Filling Phase
Blue=0 points
Yellow=+5 points
Green=+10 points
Pdet
Pdet > 0 cmH2O
< 80% of phase
Pdet > 0 cmH2O >
80% of phase
Pdet > 0 cmH2O
continuously
Zero All
Pressure
Used during
Filling
Not Possible
Not used during
Filling
Infusion Pump
Pediatric
Infused Volume
< 10% EBC
Not Possible
Infused Volume ≥
10% EBC
Infusion Pump
Adult
Infused Volume
< 10% EBC
Not Possible
Infused Volume ≥
10% EBC
Sensations
None marked
2 marked
3+ marked
Stress Test
Not Possible
Performed @ Vinf >
150 ml
Max Pves > 30
cmH2O if Leak,
>60cmH2O if no
leak
Pre-test QC
(Out of Scope)
Not performed
Performed but not
passed
Baseline & Cough
passed
Cough Check
(Out of Scope)
Not performed
Done at least once
after FS
Done every 100-150
ml infused
Voiding
Phase
Blue=0 points
Yellow=+5 points
Green=+10 points
Volume Voided
Pediatric
No uroflow
volume recorded
Volume voided > 0
Volume voided
≥50% EBC
Volume Voided
Adult
No uroflow
volume recorded
Volume voided > 0
Volume voided
>150ml
Permission to
void
Not marked
Not Possible
Marked
Pdet
Pdet > 0 cmH2O
< 80% of phase
Pdet > 0 cmH2O >
80% of phase
Pdet > 0 cmH2O
continuously
PVR
No PVR value
entered
Not Possible
Any PVR value,
including 0 ml
Table 8: QC Score Allotment Process
NOTE: Scientific references are available in the Help Menu of Synergy. Click the Help
icon available in the title bar of the main window to launch the Help Menu.