Owner's Manual

Table Of Contents
Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 76
Disinfection Devices should be disinfected in accordance with the frequency
defined below.
o Roam NXT, PIM NXT, Urocap NXT, UPP Puller NXT: Perform disinfection after
the device has contact with the patient’s skin, contact with a contaminated
catheter, or after contamination has been observed (e.g. presence of urine on
the device), or, at minimum, daily.
o NXT Go Pump Hub and PC: Perform disinfection of the device, at minimum,
once a week.
5.1.1.3 Methods and Tools
All cleaning and disinfection should be performed using moistened cloths and a soft
bristle brush.
No additional tools are required for cleaning or disinfection.
Inspect the Computer, NXT modules, and peripherals for physical damage or wear
and tear prior to cleaning.
Ensure the charging contacts for all modules are wiped free of debris and
contaminants.
Optional reusable accessories used in conjunction with the NXT System should be
cleaned and/or disinfected according to the instructions contained in their device
specific Instructions for Use. Refer to the Introduction section on page 12 an
overview of standard equipment and accessories.
CAUTION: No abrasive tools should be used. Do not spray the cleaning reagents on
the NXT System components, modules, accessories, or peripherals. Do not autoclave the NXT
System components, modules, accessories, or peripherals or clean them with sharp tools or
abrasives. Do not immerse the NXT System components, modules, accessories, or
peripherals. Do not immerse electrical connectors in water or other liquids.
Cleaning and Disinfection Preparation
To begin cleaning or disinfection preparation, complete the following:
Put on gloves and safety glasses.
Obtain the module for cleaning and select the cleaning wipes.
Put on any other specific personal protective equipment recommended by the
wipe’s manufacturer.
Ensure a stable surface is available to place any modules or parts that may need
to be separated from the NXT System during the cleaning and disinfection process
in the designated reprocessing area.
5.1.2.1 Containment and Transportation
Transport the soiled device individually in a closed container to a designated cleaning and
disinfection area.
Do not transport the device in a container with other devices.
Do not soak the device prior to cleaning.
5.1.2.2 Urocap NXT Preparation
IMPORTANT: Ensure that the plug is fully inserted to avoid fluid ingress. Refer to
Figure 88: Urocap NXT Controls.
Status DRAFT Effective