Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 79
5.1.3.3 Cleaning the NXT Go Pump Hub and the NXT Go
Infusion Hook
To clean the outer surface area of the NXT Go Pump Hub and the NXT Go Infusion Hook
follow the instructions provided in the Performing Manual Cleaning section. To clean the
Infusion Pump NXT rollers, located on the NXT Go Pump Hub (Figure 64), open the pump
door and wipe each roller with an approved cleaning reagent. Refer to the Cleaning and
Disinfection Reagents section on page 75.
Figure 64: Pump Rollers
5.1.3.4 Cleaning the UPP Puller NXT
Prior to beginning the cleaning procedure ensure to separate the UPP Puller NXT nose from
the motor. Remove the Catheter Guide and the Catheter Clamp and clean them separately.
Refer to the Assembling the UPP Puller NXT section on page 22 for a visual of how these
parts are assembled and disassembled.
To clean the UPP Puller NXT Nose, Catheter Guide, and Catheter Clamp, follow the
instructions provided in the Performing Manual Cleaning section on page 79.
If soiling is observed on the UPP Puller NXT arm or motor, wipe the surface with a fresh
cleaning wipe to remove soiling. Take care not to remove the o-ring from the motor coupler
while wiping (Figure 65).
Figure 65: Cleaning the Area Between Motor and Nose
5.1.3.5 Performing Manual Cleaning
This section covers the cleaning method for the Remote Control NXT, and peripherals.
For the Roam NXT, PIM NXT, Urocap NXT, NXT Infusion Pump, NXT Go Pump Hub and UPP
Puller NXT, ensure to refer to the device specific cleaning sections starting on page 77 for
guidance on when and where to apply the cleaning procedure described below.
Ensure to review General Cleaning and Disinfection, Cleaning and Disinfection
Preparation, Cleaning, and Completing Cleaning and Disinfection sections fully
before beginning cleaning.
Turn off and disconnect all power to the NXT System or NXT module.