Owner's Manual

Table Of Contents
Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 82
To use the system after cleaning or disinfection, re-connect power to the system, dock
devices, and follow the instructions provided for normal use.
5.1.5.2 Packaging and Storage
Devices may be packaged flat in a clean, anti-static box or wrapped in a lint free cloth post-
cleaning or disinfection. Do not stack devices on top of each other and ensure that devices
are not pressed against each other or against the container. If the box has multiple
units, use a divider to ensure no contact with other devices.
Refer to the Appendix on page 110 for Storage Conditions.
For long term storage, LABORIE recommends that the Roam NXT and the Urocap NXT be
fully charged prior to storing. Press the “Reset” button on the module to cause the device
to go into long-term sleep mode. To wake up the modules after storage, pair them with the
laptop PC. Refer to the Hardware section on page 96.
5.2 Preventive Maintenance
General Maintenance
This section provides instructions on maintenance activities which can be performed by the
user on various modules of the NXT System. The following sections, beginning on page 82,
should be referenced for device specific information.
Caring for the Interlocked Roam NXT
Caring for the Urocap NXT
NOTE: Module batteries are not user replaceable, except for the Remote Control NXT.
Contact LABORIE Service for battery replacement if required. Refer to the Hardware section
on page 96 for an equipment overview and instructions on charging batteries.
5.2.1.1 Caring for the Interlocked Roam NXT & PIM NXT
Inspect the Roam NXT and PIM NXT to ensure there are no signs of visible deterioration of
the device prior to using them for a Urodynamics Study. General guidelines for caring for
these devices are provided below:
If a transducer door or belt clip breaks, is damaged or lost, contact LABORIE
Service for a replacement.
EMG recording cables should be replaced either at sign of wear or annually.
The PIM NXT should be replaced every two years to sustain recording quality.
As a part of regular maintenance, inspect the PIM NXT for T-DOC transducer
channels for any obstructions, dirt, or debris. Any objects inserted into the PIM
NXT transducer channels other than the intended catheter connector may cause
damage and channel obstruction.
If an object is trapped inside the cavity, do not insert a catheter into that channel
until the object is removed.
If small objects or an O-ring get stuck inside the charging cavity, turn the cavity
upside down and gently shake the PIM NXT to remove the object.
Do not attempt to remove any lodged objects from the PIM NXT for T-DOC
transducer channels with a tool. Do not attempt to clean or remove dirt or debris
from the transducer channels. Contact LABORIE Service for repair or replacement.
Status DRAFT Effective