5.3.3.1 Checking the Total Weight without a Saline Bag
To begin calibration verification, measure the total weight of the Infusion Pump NXT and
Infusion Volume Transducer without a saline bag:
a) Setup the Infusion Volume Transducer without a saline bag.
b) Click the Settings icon in the title bar, then click System Settings > Device
Manager option. The Device Manager window will pop-up.
c) Click the Infusion Pump NXT option, then click the Run Diagnostics button. The
device diagnostics window will pop-up (Figure 70).
d) Press the Zero button.
e) The Total Weight should measure as zero immediately after zeroing. To complete
calibration verification, check the total weight with a saline bag.
Figure 70: Infusion Pump NXT Diagnostics Window
5.3.3.2 Checking the Total Weight with a Saline Bag
To complete calibration verification, check the total weight of a 500mL saline bag:
a) Press the Zero button.
b) Measure the weight of the saline bag using an external measuring device.
c) Setup the Infusion Volume Transducer with the saline bag.
d) The Total Weight in the diagnostic window should be the previously measured weight
+/- 5%.