Owner's Manual
Table Of Contents
- 1 Introduction
- 2 Quick Start
- 2.1 Setting Up the Equipment
- 2.2 Gathering Supplies and Equipment
- 2.3 Preparing the Urocap NXT
- 2.4 Preparing the NXT Go Pump Hub
- 2.5 Preparing the Roam NXT
- 2.6 Preparing the UPP Puller NXT (Optional)
- 2.7 Setting Up Consumables and Preparing the Patient
- 3 How to Perform a Study
- 4 Synergy Software
- 5 Maintenance and Service
- 5.1 Cleaning and Disinfection
- 5.2 Preventive Maintenance
- 5.3 Verifying Calibration
- 5.4 Configuration
- 5.5 Computer Virus Protection
- 5.6 Disposing of Product After Use
- 5.7 Environmental Consideration of Waste Disposal
- 5.8 Backups
- 5.9 Troubleshooting Guide
- 6 Hardware
- 7 Appendix
![](/manual/laborie-medical-technologies/tdoc3/owner-s-manual-english/images/img-90.png)
Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 89
5.4 Configuration
NXT Go Electrical Isolation Diagram
Refer to the diagram shown below in Figure 71 for a visual representation of the NXT Go
electrical isolation from mains supply. For further information regarding electrical isolation,
contact LABORIE Service.
Urocap NXT
Direct electrical contact
Indirect electrical contact
100-240VAC
50/60Hz
100-240VAC
50/60Hz
Isolation Barrier1
Isolation Barrier2
Isolation Barrier2
UPP Puller NXT
PIM1
Roam Base 1
Infusion Fluid
Pressure
EMG
Figure 71: NXT Go Isolation Diagram
Roam NXT and Wall Charger Isolation Diagram
Refer to the diagram shown below in Figure 72 for a visual representation of the method to
isolate the Roam NXT from supply mains when configured with a wall charger. For further
information regarding electrical isolation, contact LABORIE Service.
Roam
Direct electrical contact
Indirect electrical contact
100-240VAC
50/60Hz
Isolation Barrier1
NXT Module Power
Supply
P/N:POW1033
Cable NXT POGO Pin to
BINDER 3 meter
P/N:CAB1081
Isolation Barrier2
PIM
EMG
Pressure
Figure 72: Roam NXT and Wall Charger Isolation Diagram
Status DRAFT Effective