Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 92
Symptoms
Possible Cause(s)
Corrective Action(s)
Devices not connected in
software?
Reconnect the devices to the system
through the USB cables provided.
Consult Device Manager Window to
confirm device connection.
Computer or Printer
cannot power on?
No power on electrical
outlet?
Ensure electrical outlet is working,
power cords are not damaged, and all
power cords are secure at both ends.
Damaged power cord?
Power cord not
connected properly?
EMG
EMG reading too
high/low?
EMG channel scale not
optimized?
Set EMG scale in the Workflow
Configuration for Pressure flow.
Electrodes are wet or
generally not sticking?
Shave area if necessary and wipe dry.
Apply a good amount of tape to keep
moisture out.
Detached electrode?
Re-attach electrode.
Water/urine leaked over
the electrodes?
Dry the area with a towel and replace
the electrodes.
No spike response
in EMG and
baseline is flat?
Electrodes are not
picking up the proper
muscle group?
Move either one or both measuring
electrodes (the ones attached to the
red & black leads) closer together.
EMG response is
too high and
channel saturated?
Electrodes not picking up
the proper muscle
group?
Move either one or both measuring
electrodes (the ones attached to the
red & black leads) closer together.
EMG reading is too
high or channel
saturated?
Mains frequency pickup?
Check the measurement electrodes and
quality of patient ground.
Pressure
Pressure not
responding?
Position of catheters?
Check position and adjust as
necessary.
Catheters are not
securely connected to
the PIM NXT?
Check connections and adjust as
necessary.
Status DRAFT Effective