Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030 93
Symptoms
Possible Cause(s)
Corrective Action(s)
Zeroes not set properly?
Reset zeroes in Preview workflow step.
Catheter kinked?
Replace catheter if necessary.
Pump
Pump not running?
Pump pressure limit
reached?
Reset Pves to zero. Check for kinked
bladder catheter. Ensure Pump
pressure does not exceed 150 ml/min.
Pump settings?
Check Pump and Infusion lines. Check
if Pump door is closed. Check if Pump
tubing is inserted correctly.
Urocap
Uroflowmeter signal
shows vibration
and/or spike
patterns?
Plastic beaker is touching
the flow funnel?
Reposition the Uroflowmeter and re-
check.
Patient has touched the
Uroflowmeter with their
feet?
Ask patient to remain calm during
procedure.
Floor is unsteady?
Move to a more solid foundation.
Urocap NXT is recording
environmental pressure
changes?
Unplug the vent plug from the Urocap
NXT air vent located on the bottom of
the unit.
Activate the Urocap Noise Reduction
Level. Click Settings > Workflow >
Editor to launch the Workflow
Configuration window then, select
the Global Tab. The Urocap Noise
Reduction Level feature provides
three options: None, Narrow, or
Wide. Consult LABORIE Service for
guidance on setting selection.
Incorrect flow or
volume readings?
Beaker is not properly
seated on the
Uroflowmeter dish?
Adjust beaker position to ensure
beaker is placed flat in the center of
the Uroflowmeter.
Beaker placed or
adjusted during study?
Ensure that the beaker is placed on the
Uroflowmeter before starting the study.
Funnel is touching the
beaker?
Adjust commode chair or
Uroflowmeter.