MAINTENANCE
Attaching the Right Arm
Open the torso as described in steps 1 - 4, Opening the Torso.
Follow the procedures listed below in reverse to detach the arms.
1 Align the arm axle with the shoulder socket.
2 Feed the cables from the arm axle through the shoulder socket
3 Push the axle into the bracket until it is flush with the inside of
the bracket.
4 Tighten the shoulder screw with the Allen wrench.
JAW ANT RIGHT
ARM
LA CONV
JAW
AN
5 Connect the corresponding cables as shown below.
Right Arm to Torso – Cable and Tube Descriptions
Name/Label Tube/Cable Color Connector Description
Right Arm Black harness cable Black rectangular
connector, 4 lead
Jaw Ant Coax Round silver connector
To LA Conv Black, insulation Black rectangular
connector, 8 lead
Replacing IV Arm Catheter/Filter
Replace the IV catheter module if it is damaged or if resistance is
unrealistically high when injecting fluids.
1 Unzip and fold down the IV arm skin to expose the IV catheter
module
2 Using the no. 4 Phillips screwdriver; remove the 4 screws and
pull the IV module out of the arm.
3 Lift the IV catheter up and out, exposing the recess where the
filter is located.
4 Lift out the IV filter and replace with a new filter.