TROUBLESHOOTING
Simulator> dialog on LLEAP will indicate which Patient Simulator
is pulse palpated.
Airway Contamination
Problem
Simulator airways have become contaminated from mouth-to-
mouth rescue breathing
Solution
Clean the outside of the Patient Simulator with manikin wipes.
Clean the inside of the oral cavity with manikin wipes. Change
the lung bladders; see: Maintenance - Replacing Lung Bladders.
Note: The Patient Simulator airways are not designed for
mouth to mouth rescue breathing or to be disinfected.
Chest Movement
Problem
No chest rise on Patient Simulator.
Solution
Check that Patient Simulator is switched ON.
Check that Patient Simulator is not in sleep mode due to
inactivity. Reactivate the Patient Simulator.
Check that awRR is not set to zero on Instructor Application
(Patient Case or scenario)
Check that airway complications like maximum air resistance or
laryngospasm are not set.
Check if the internal compressor is switched off. See: Turning the
Internal Compressor Off/On.
Check that any external compressed air source is switched off
and that the air tube is disconnected from the Patient Simulator.
The internal compressor may have overheated. Wait
approximately 20 minutes for it to cool down. Remove blankets
or covers from the Patient Simulator, open the torso skin to
facilitate cooling.
Chest-rise is set to bilateral (for example if ET-tube is inserted
too far into the bronchia).
Chest-rise bladder is leaking or tubing to chest-rise bladder is
twisted, kinked or disconnected. Replace chest-rise bladder if it
is leaking, see Maintenance: Replacing Chest Rise Bladders.
Check air tubing for leakage; check that all connections are intact.
Replacing tubing if necessary.
Shallow chest movements and the internal compressor runs
continuously. The internal compressor may be worn - contact
Laerdal Technical Service.
Blood System
Problem
No blood flow.
Solution
Make sure that the internal reservoir is filled with fluid.
Problem
Too low flow.
Solution
Clean the blood system.
Check the flow setting in the Instructor Application.
The blood system may need to be re-calibrated. Contact your
local service representative.
Replace filter in pelvis.
Problem
Air when bleeding
Solution
Bleed the blood reservoir empty, then refill.
Problem
Simulated blood is leaking from the back of the right leg during
filling or when the Patient Simulator is turned off.
Solution
The internal reservoir may need to be replaced. Contact your
local service representative.
Problem
Can’t fill the blood reservoir.
Solution
The filter inside the fill bottle may be clogged. If so, order a new
fill bottle
Simulator Limbs
Problems
Lack of motion in the legs.
Possible Solution
Loosen and re-adjust the hip joint nuts on the inside of the
pelvis. See Maintenance for details on how to attach the legs