Manual

44
Lungs
Problem
Lungs not functioning properly.
Possible Solution
Check that the airway resistance is not set to maximum in the
Instructor Application.
Open the torso and chest plate. Check that the lungs are free to
expand and are not restricted by any cables.
Check that the lung bladders are properly connected, and that
the tubes are not twisted.
Check that the lung bladders are in a horizontal position and
inserted correctly. Ensure that the lung compliance O-rings lie
between the folds of the lung bladders.
Check for flaws or ruptures in the lung bladders.
Check that the two lung compliance O-rings are fitted correctly.
Replace O-rings if they appear to be damaged.
Check that there are no obstructions inside the simulator
airways which may block air flow.
If there is no change when adjusting lung compliance, contact
Laerdal Technical Service.
If there is no change in lung resistance, contact Laerdal Technical
Service.
Mechanical Noise during Auscultation
In the Instructor Application, click <auscultation focus>.
Batteries
Problem
Battery life is less than 150 minutes with healthy patient and fully
charged batteries.
Possible Solutions
Batteries may be old, (recommended lifetime is 200 discharge
cycles). Insert new batteries.
Compressor may not function properly; consult your local
Laerdal Technical Services Centre.
Pulses
Problem
Cannot feel pedal pulses
Possible Solutions
Skin may be too tight over pulse units – readjust skin and reboot.
Clinical Features - Pneumothorax Bladders
Problem
Experience problems with pneumothorax bladders.
Solution
Check tubing connection at base of the bladder, to ensure that
tubing has not been disconnected.
Simulator Shutdown
Problem
The Patient Simulator is unresponsive.
Solution
Press and hold the <ON/OFF> button for 10 seconds to force the
Patient Simulator to shutdown.
TROUBLESHOOTING