Temperature Coefficient
The temperature coefficient is the amount of change in
conductivity per degree of temperature; it is expressed in
percent per °C. Entering the exact temperature
coefficient of the solution being measured will accurately
compensate the temperature for almost any solution. The
temperature coefficient can be set between 0.0 and 3.0 %
per °C. The meter default is 2.1% per °C.
6. Enter Advanced Set-up.
7. Press or button until [t.Co %] appears on the
LCD. Press ENTER button.
8. Press or button to select a value between 0.0 and
3.0.
9. Press ENTER button to select. The meter will return
to the menu, [t.Co %].
Press or button to move to the next menu or press CAL to exit to the
measurement mode.
Normalization Temperature
The meter can be set to normalize the conductivity measurements to a
standard temperature of either 25 °C or 20 °C.
The default value is 25 °C.
1. Enter Advanced Set-up.
2. Press or button until [t.nr °C] appears on the
LCD. Press ENTER button.
3. Press or button to select either [25.0 °C]or[20.0
°C].
4. Press ENTER button to select. The meter will return
to the menu, [t.nr °C].
5. Press or button to move to the next menu or
press CAL to exit to the measurement mode.
Press ENTER to confirm selection; or
Press CAL to escape without confirming selection
%
%
%
HOLD
ENTER
Press ENTER to confirm selection; or
Press CAL to escape without confirming selection
C°
C°
C°
HOLD
ENTER
27