INTERFERENCES: The following substances produce a positive interference at
the concentration given:
Chloride (Cl
–
) 7000 mg/L
Phosphate (PO
4
–3
)16mg/L
Hexametaphosphate (NaPO
3
)
6
1mg/L
The following substances produce a negative interference at
the concentration given:
Alkalinity (CaCO
3
) 5000 mg/L
Aluminum (Al
3+
)0.1mg/L
Iron (Fe
3+
)10mg/L
Sulfate (SO
4
–2
) 200 mg/L
Color and turbidity must be removed or compensated for in
the procedure. Temperature should be maintained within
5°C of room temperature.
Fluoride 1/3 SMART SPECTRO TEST PROCEDURES 05.04