Owner's manual

PROCEDURE
Use universal sample holder.
1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing
90 Sul de-LR) from TESTING MENU.
4. Scroll to and select 90 Sul de-LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro. Use the 1.0 mL pipet (0354) to add 1.0 mL of
*Sul de Reagent A (V-4458). Cap and mix.
8. Add 6 drops of Sul de Reagent B (V-4459). Cap and mix. Wait 1 minute.
Solution will turn blue if sul des are present.
9. Use the 1.0 mL pipet (0354) to add 2.0 mL of Sul de Reagent C (4460). Cap
and mix. Color development is immediate and stable.
10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record
result.
11. Press OFF button to turn spectrophotometer off or press EXIT button to exit to
a previous menu or make another menu selection.
SMART Spectro Test Procedures 2.11 SULFIDE
Test Procedures