Owner's manual
6 SMART Spectro Operator’s Manual 2.11
SPECIFICATIONS
INSTRUMENT TYPE: Single beam spectrophotometer
Readout 5 line, 18 character per line LCD
Wavelengths 350-1000 nm
Wavelength Accuracy ± 2 nm
Wavelenth Resolution 1 nm
Wavelength Bandwidth 5 nm (max)
Photometric Range 0-125%T, –0.1-2.5A
Photometric Accuracy ± 0.005A
Photometric Stray Light <0.5%T
Dispersive Device 1200 Lines/mm ruled grating
Sample Chamber Accepts 25 mm diameter fl at-bottomed test tubes,
10 mm square cuvettes, 16 mm COD test tubes
Source Lamp Quartz halogen
Modes %/T, ABS, pre-programmed tests
Pre-Programmed Tests YES, with automatic wavelength selection
Languages English or Chinese
User Defi ned Tests Up to 25 user tests can be input
RS232 Port 8 pin mDIN
Power Requirements Battery Operation (optional): Ni-Metal Hydride
battery pack Line Operation: 110/220V, 50/60 Hz
Dimensions 36 cm (wide) x 28 cm (deep) x 17 cm (tall)
Weight 10.3 lbs, 4.65 kgs
STATISTICAL AND TECHNICAL DEFINITIONS
RELATED TO PRODUCT SPECIFICATIONS
Method Detection Limit (MDL): “The method detection limit (MDL) is defi ned as
the minimum concentration of a substance that can be measured and reported
with 99% confi dence that the analyte concentration is greater than zero and is
determined from analysis of a sample in a given matrix containing the analyte.”
1
Note that, “As Dr. William Horwitz once stated, ‘In almost all cases when
dealing with a limit of detection or limit of determination, the primary purpose of
determining that limit is to stay away from it.’”
2
1. CFR 40, part 136, appendix B
2. Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and
L Vanatta, American Laboratory, Sept 2003, P. 31.