Chapter 3 - American GPR Index Cover & GPR Install Landis+Gyr
34 98-1135 Rev AB Installation Guide
Figure 3 - 7. Remove Index Cover Tamper Seals and Screws
2. Gently push the Badger dial wheel toward the front of the index cover and away from the index
dial pointer. Lift the index out of the cover.
Figure 3 - 8. Disengage Badger Dial Wheel from Index; Remove Index from Cover
3. Inspect the index and replace it if any of the following conditions exists:
Pointers are loose on their shafts.
Index face contains cracked or peeling enamel.
Index drive mechanism does not rotate easily.
U WARNING:
The Index Face Mount Dial Wheel (40-1742) must be used with metal pointer indexes.
The Balanced dial wheel (40-1538) must be used on plastic pointer indexes.