Landis+Gyr Chapter 4 - Rockwell GPR Index Cover & GPR Install
Installation Guide 98-1135 Rev AB 49
Figure 4 - 6. Commercial Module Cover Removed
4. Swipe each end of the module with an RF Buster or magnet to force the module to save meter
data to non-volatile memory. Remove the Commercial Gas Module tamper seals, screws, and
cover. Disconnect the battery.
5. The Commercial Gas Module must be disassociated from the meter.
A. If RIMS is used to manage gas module installations, leave the battery disconnected, replace
the module cover, and return the module to the Landis+Gyr cross-dock.
B. If GPrep is used to install the gas modules, connect the laptop PC and Shooter Box or USB
Shooter programming cable to the module Serial/Power connector. Follow the procedures
on using GPrep “MRB Mode” outlined in the “GPrep User Guide,” Landis+Gyr publication
98-1119.
6. Disconnect the module from the programming device. Leave the battery disconnected, replace
the module cover, and return the module to the Landis+Gyr cross-dock.
Rockwell Mini Switch Kit Installation
Meter Preparation
1. Inspect the index and replace it if any of the following conditions exists:
Pointers are loose on their shafts.
Index face contains cracked or peeling enamel.
Index drive mechanism does not rotate easily.
NOTE: The GPR module gets readings from a dial wheel mounted on the vertical drive shaft behind
the index. The cover is installed on the meter with the sensor in the cover positioned on the BACK
side of the index.
Module Cover Tamper Seal
Location
Serial/Power
Connector