Table 18 Handheld Fluorescence Camera Sterilization Instructions
3.
Before sterilizing, ensure the device has
been properly cleaned according and that
the USB-c hatch is closed and secured in
place.
4.
In the non-sterile area of the dedicated workspace, prepare a
basin containing liquid chemical sterilant according the
manufacturers instructions to achieve the appropriate dilution
5.
Ensure the HFC is cleaned and completely dry prior to immersing in the sterilant.
6.
Fully immerse the HFC in the liquid chemical bath ensuring all
surfaces of the device are fully immersed.
7.
Soak the device for 10 hours, monitoring (e.g. temperature) as
per the sterilant manufacturer’s recommendations
8.
Remove the HFC using sterile technique and rinse according to the sterilant manufacturer’s
instructions.