Use Instructions

PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 58 of 101
5.4.2. Disinfecting CCD
Do not continue with disinfecting the CCD, the CCD Power Supply, or cables unless the cleaning process
described in Section 5.4.1 is completed. Effective disinfection can only be achieved after completing the
cleaning process.
Table 24 Contact Charging Device Disinfection Instructions
1.
After cleaning, use at least two new Intermediate Level Disinfection Wipes to thoroughly wet ALL
SIDES and surfaces of the CCD, the CCD Power Supply, and Cables.
2.
Thoroughly wipe crevices and areas that are difficult to reach. Where required, fold or bunch
Intermediate Level Disinfection Wipes to push into crevices, corners, and areas that are difficult to
reach.
3.
Use additional Intermediate Level Disinfection Wipes as necessary to ensure ALL SIDES and surfaces
of the CCD, the CCD Power Supply, and Cables have been wet.
4.
Once wet, allow the CCD, the CCD Power Supply, and Cables to dry.
5.
The CCD, the CCD Power Supply, and Cables are now disinfected and ready for use. Observe sterile
technique when transporting equipment from re-processing locations to points of use.
5.5. PSC Protective Storage Cradle Cleaning and Sterilization
Before use, clean and sterilize the Protective Storage Cradle.
5.5.1. Cleaning Protective Storage Cradle
Table 25 Protective Storage Cradle Manual Cleaning Instructions
1.
Prepare Prolystica 2x Concentrate Presoak and Cleaner enzymatic detergent solution per
manufacturer’s instructions (1/8 fl. oz./gal to 1/2 fl. oz./gal) with lukewarm tap water (22°C to 30°C
temperature range).
2.
Immerse the device(s) in the prepared solution and soak for one (1) to five (5) minutes.
3.
Use a non-linting wipe or cloth saturated with the detergent solution to wipe all device surfaces until
visibly clean.
4.
Rinse the device(s) in a tap water (or better) bath for at least 30 seconds.
5.
Thoroughly dry all surfaces of the device(s) using non-linting wipes or cloths, changing wipes/ cloths
when necessary to ensure the devices are completely dry.
6.
Visually inspect each device for the absence or presence of remaining soil in a well-lit area. If soil is
present, repeat the manual cleaning steps until all visible soil is removed.