PN 10232 Eagle V1.2 Instructions for Use Rev 1.0 Page 81 of 101
Table 39 Custom Sterile Sleeve Wireless Configuration Instructions
14.
Pull CSS slack material taut from
the bottom and left, and wrap the
HFC in a gift-wrap style.
15.
Pull off the adhesive backing
from the CSS Tape while keeping
the gift-wrapped CSS material in
position.
Lay the CSS Tape over the gift-
wrapped CSS material.
16.
Lay and push the CSS Tape firmly
across the backside of the HFC
body.
Flip the HFC around and finish
adhering the CSS Tape around
the right front side of the HFC.
17.
The Handheld Fluorescence Camera is now sheathed in the Custom Sterile Sleeve using the ECH Wireless
Configuration (Section 6.3.2.1), and ready for Imaging per Section 6.2.2.2.