Figure 10 Setting up version 2 of the wireless unit for reforming
2 Follow the timer instructions to program the on/off
cycles listed in Table 1.
Set the timer to four cycles of 5 minutes on and 1 minute
off, then three cycles of 15 minutes on and 3 minutes off,
then six cycles of 30 minutes on and 10 minutes off. The
final cycle (14) is set to on with no off. Allow the unit to
charge overnight (at least 8 hours) in the final cycle.
Cleaning the Wireless Unit
Clean the wireless unit if dirt or blood is present, or as defined
by your institution’s infection control policies.
1 Clean the exterior of the unit, including the seams
along the battery door, using a cloth moistened with a
10% bleach solution. Follow with a cloth moistened
with water to remove residual bleach.
NOTE: For a complete list of approved cleaning agents,
refer to the Specifications section of the
SureStepFlexx Meter Operator’s Guide.
2 Dry the unit thoroughly.
CAUTION: Do Not get water inside the wireless unit.
Never immerse the unit or hold it under running water.
Tab le 1: Charge/Discharge
Times for 14 Cycles
Cycle On/Off Duration
Cycle 1
5 min / 1 min
Cycle 2
5 min / 1 min
Cycle 3
5 min / 1 min
Cycle 4
5 min / 1 min
Cycle 5 15 min / 3 min
Cycle 6 15 min / 3 min
Cycle 7 15 min / 3 min
Cycle 8 30 min / 10 min
Cycle 9
30 min / 10 min
Cycle 10
30 min / 10 min
Cycle 11
30 min / 10 min
Cycle 12
30 min / 10 min
Cycle 13
30 min / 10 min
Cycle 14 charge overnight
ethernet.book Page 22 Tuesday, December 11, 2007 11:25 AM