Use Instructions

28
Operating the Device
en
5.10 Suction
The suction system consists of the following:
•Vacuum tube
Hygiene filter
Suction container
•Drain tube
Standard instrument or suction/irrigation instrument
1. Connect device with the suction container (using vacuum tube and hygiene
filter). The vacuum tube set with hygiene filter can remain connected to the
device for max. 30 days.
2. Connect suction container with instrument (using drain tube).
Close stopcock if using instrument with stopcock.
3. Press the suction start/stop key (LED lights up).
The device generates a negative pressure inside of the suction system. The
vacuum pump stops when a negative pressure of max. -60 kpa (-0.60 bar) has
been reached.
The vacuum pump automatically reactivates, if the min. negative pressure is
reached.
4. Starting suction:
Actuate the trumpet valve or open the stopcock.
5. Stopping suction:
Release the trumpet valve or open the stopcock.
6. Press the suction start/stop key again (LED goes out).
CAUTION!
The full suction output/volume is available only, if the suction system is pre-
evacuated (vacuum generation). Pre-evacuation takes about 20 to 60 seconds
depending on the volume of the secretion container.
NOTE!
Only use suction containers with overflow protection.
Fig. 5-14 Suction system with suction-/
Irrigation instrument
(1) Vacuum tube set with hygiene fil-
ter
(2) Suction container
(3) Suction tube
(4) Irrigation tube
(5) Instrument
(1)
(2)
(5)
(4)
(3)
Fig. 5-15 Suction system with standard
instrument
(1) Vacuum tube set with hygiene fil-
ter
(2) Suction container
(3) Suction tube
(4) Standard instrument
(1)
(2)
(3) (4)
1./2.