Data Sheet

EL-USB-2-LCD
Temperature, Humidity and Dew Point
Data Logger with LCD Screen
www.lascarelectronics.com
Issue 11 08/2016 N.R Page 1 of 5
This standalone data logger measures and stores over 16,000 temperature and humidity readings from -35 to +80°C (-31 to +176°F) and
0 to 100%RH range at a resoluon of 0.5°C (1°F) and 0.5%RH.
The user can easily set up the logger and view downloaded data by plugging the data logger into a PCs USB port and using the free
EasyLog soware. Data, including calculated dew point, can then be graphed, printed and exported to other applicaons for detailed
analysis.
The high contrast LCD can show a variety of temperature and humidity informaon. At the touch of a buon, the user can cycle between
the current temperature and humidity, along with the maximum and minimum stored values for temperature and humidity.
The data logger is supplied with a lithium metal baery, giving up to 2 years’ logging life. The logger is protected against ingress from
water and dust to IP67 standard when the cap is ed.
▪ -35 to +80°C (-31 to +176°F) and 0 to 100%RH measurement range
▪ Stores over 16,000 readings for both temperature and humidity
▪ EasyLog soware available as a free download
▪ High contrast LCD, with two and a half digit temperature and humidity
display funcon
▪ Logging rates between 10 seconds and 12 hours
▪ Immediate, delayed and push-to-start logging
▪ User-programmable alarm thresholds for both temperature and humidity
▪ Status indicaon via red/green LEDs
▪ Environmental protecon to IP67
Temperature Measurement range -35°C to 80°C (-31°F to 176°F)
Internal resoluon 0.5°C (1°F)
Accuracy (overall error)* 0.55°C (1.04°F) typical (5 to 60°C)
Long term stability <0.02°C (0.04°F) / year
Relave Humidity Measurement range 0 to 100%RH
Internal resoluon 0.5%RH
Accuracy (overall error)* 2.25%RH typical (20 to 80%RH)
Long term stability <0.25%RH / year
Dew Point Accuracy (overall error)* 1.7°C typical
(-35 to 80°C, 40 to 100%RH)
Logging rate User selectable between 10
seconds & 12 hours
Operang temperature range -35 to +80°C (-31 to +176°F)
Baery life 2 years (at 25°C and 1 minute
logging rate, LCD on)
Readings 16,382 temperature, 16,382
relave humidity
Dimensions 126 x 25 x 22mm
(4.96 x 0.98 x 0.86”)
* The overall error takes in to account the sensor accuracy (as shown on page 4) and the resoluon of the
data logger
SPECIFICATIONS ACCESSORIES
Replacement baery
Handheld data logger
programmer & collector
BAT 3V6 1/2AA
EL-DataPad
INCLUDED IN THE BOX
Baery
Mounng Bracket
BAT 3V6 1/2AA
EL-LCD WALL BRACKET
CALIBRATION CERTIFICATES NOW AVAILABLE
Lascar now oers a Traceable Calibraon Cercate Service on Temperature Data Loggers. Using reference equipment
which has been calibrated by a UKAS/NIST accredited laboratory and using apparatus traceable to naonal or
internaonal standards. For more informaon, please see www.lascarelectronics.com.