Data Sheet

This standalone data logger measures and stores up to 16,382 relativehumidity and 16,382 temperature readings over 0 to 100%RH and -35
to +80°C (-31 to +176°F) measurement ranges. The user can easily set up the logging rate and start-time,
and download the stored data by plugging the data logger into a PC’s USB port and running the purpose designed software under Windows
2000, XP and Vista (32-bit). Relative humidity, temperature and dew point (the temperature at which water vapor present in the air begins to
condense) data can then be graphed, printed and exported to other applications. The data logger is supplied complete with a long-life
lithium battery, which can typically allow logging for a year. Status indication is via ashing red and green LEDs. The logger is protected
against ingress from water and dust to IP67 standard when the plastic cap and seal are tted.
EL-USB-2
Humidity, Temperature and
Dew Point USB Data Logger
Specications liable to change without prior warning Issue 10 07/2012 S.L Applies to EL-USB-2 Page 1 of 4
FEATURES
• 0 to 100%RH measurement range
• -35 to +80°C (-31 to +176°F) measurement range
• Dew point indication via Windows control software
• USB interface for set-up and data download
• User-programmable alarm thresholds for %RH & T
• Status indication via red and green LEDs
• Supplied with replaceable internal lithium battery
and Windows control software
• Environmental protection to IP67
ORDERING INFORMATION
Standard Data Logger
(Data Logger, Software on
CD, Battery)
Replacement Battery
EL-USB-2
BAT 3V6 1/2AA
www.lascarelectronics.com
* This species the overall error in the logged readings, for relative humidity measurements between 20 and 80%RH.
** This species the overall error in the calculated dew point, for relative humidity measurements between 40 and 100%RH at 25°C.
*** Depending on sample rate, ambient temperature and use of alarm LEDs.
Specications Minimum Typical Maximum Unit
Relative Humidity
Measurement range 0 100 %RH
Repeatability (short term) ±0.1 %RH
Accuracy (overall error) ±3.0* ±5.0 %RH
Internal resolution 0.5 %RH
Long term stability 0.5 %RH/yr
Temperature
Measurement range -35 (-31) +80 (+176) °C (°F)
Repeatability ±0.1 (±0.2) °C (°F)
Accuracy (overall error) ±0.5 (±1) ±2 (±4) °C (°F)
Internal resolution 0.5 (1) °C (°F)
Dew Point Accuracy (overall error) ±1.1 (±2)** °C (°F)
Logging rate every 10s every 12hr -
Operating temperature range -35 (-31) +80 (+176) °C (°F)
1/2AA 3.6V Lithium Battery Life 1*** Year