EASYLOG SOFTWARE GUIDE
Conguration Completed
After the computer has configured the data logger with your chosen parameters, the following confirmation
screen is displayed.
Remove the data logger from the computer, re-fit the translucent cap, and place the data logger in the
location where you want to measure.
With the data logger connected to the PC, you can at any time check the current status of the data logger
by selecting Options > Current Status from the menu.
Note: If you leave a temperature or humidity data logger connected to the PC, then the it will operate at
an elevated temperature due to a self heating effect. This may cause the data logger to record incorrect
readings.
If the data logger is connected to the USB port for a long time (e.g. 15 minutes), then you should allow time
for the data logger to cool down after disconnecting it from the USB port.
When a data logger is connected to a USB port, the battery inside the data logger is discharged at a higher
rate than normal. To conserve battery life, do not leave the data logger connected to a USB port for a