Pay attention to the polarity information on the thermocou-
ple as well as on the device connection.
!
ThermoMaster
07
GB
1
1 x 9V Typ (NEDA 1604, IEC 6F22)
1.
2.
3.
4.
If the thermometer is subject to large × uctuations in the ambient temperature, after the temperature
has stabilised, wait 20 minutes before performing a measurement.
Always ensure adequate thermal coupling at the measuring point in order to avoid measuring errors
caused by temperature loss.
Please note that all thermometers with a contact sensor in× uence the measurement and their thermal
capacity can reduce the actual temperature. More thermal energy should therefore be applied to the
thermocouple than it can dissipate.
OL is shown in the display when no measuring sensor is connected.
– The thermometer shows OL when the measured temperature is outside the measuring range.
Only use the correct types of thermocouple plugged in at the K-type connection; the wrong type of
thermocouple can cause considerable measuring errors.
In addition to being subject to ageing, a thermocouple is also greatly depending on the operating condi-
tions and should therefore be checked regularly.
High pressure or mechanical deformation can change the grid structure and therefore have an in× uence
on the thermoelectric voltage output.