Continuous and discontinuous
A: Click On and “Off in “Continuousat right
information column.
* This function is effective only in the case of
Pattern2” and Pattern3”.
* If continuous/discontinuous is set as "On", trigger
is effected only Pattern2 and Pattern3 match
continuously Pattern state, without mixture
of other states; otherwise no continuous data is
required for triggering.
Trigger Counter
A: Enter the intended times in the entry box of
Trig Counterat right-hand information column,
within the range of 1~255 times.
* If the entry exceeds 255 times, 255 times is limited
during system setting.
Internal/External trigger
* If setting is set as " Internal" , it selects the internal
for sampling frequency.
* If setting is set as "External" , it selects the external
CLOCK for sampling frequency.
*Caution!
Use the last channel (CH31) for LA Series to be
external clock input.
Use the individual external clock input for LA
Series.