LA-series Operating Manual
LA-series Operating Manual Table of Contents Chapter 1 Product Description.............................................................01 1-1 1-2 1-3 1-4 1-5 Product Overview Safety Notes Product Specifications and Characteristics Accessory List Optional Components Chapter 2 Assembly.................................................................
LA-series Operating Manual Chapter 5 Operation of PC-LINK...................................44 5-1 Installation of PC-LINK 5-2 Operation of PC-LINK 5-3 Software System Requirements 5-4 Software Installation 5-5 Hardware Installation 5-6 Software Execution 5-7 Interface 5-8 Detailed Operations 5-9 Shortcut Flow Process Chapter 6 Calibration...............................................................94 6-1 Calibration Mode Chapter 7 Maintenance, Repair and Utilization........................
LA-series Operating Manual Warranty Description of Logic Analyzer Leaptronix provides a one-year warranty of spare parts and assembly, from the date of delivery, for the products produced and sold by itself. In the event of any defect within the warranty period, Leaptronix will deliver materials and repair the defect products at its own cost, with exception of consumables.
LA-series Operating Manual Warranty Description of Consumables of Logic Analyzer The spare parts or assembly components other than the analyzer body and junction box are consumables not covered in the warranty service. The spare parts and assembly components shall be checked within 30 days after procurement; in the case of any abnormality, the designated agents shall be notified immediately, and defect spare parts or components shall be attached for replacement.
LA-series Operating Manual Chapter 1 Product Description 1-1 Product Overview Since its founding in 1980, Leaptronix was always devoted to IC burning and measurement, and also committed itself to delivering perfect R&D environment and top-quality measuring instruments.
LA-series Operating Manual 1-2 Safety Notes All operations, maintenance and servicing must adhere to the following safety notes and precautions. Our company shall not assume any responsibility for any unexpected results arising from misuse of the instruments due to failure of following the safety notes. 1. Please make sure that input AC voltage is correct, the fuse is properly installed, and the standard specifications are noted in the Product Description. 2.
LA-series Operating Manual 1-3 Product Specifications and Characteristics Item Time sequence analysis(capture frequency) State analysis(External clock) Bandwidth Channel Memory Trigger Reference voltage Max.
LA-series Operating Manual 1-4 Accessory List NO 01 02 03 04 05 06 07 Name LA-2050 Signal capture box Signal capture cable Signal capture hook Optical disc Quantity 1 4(32CH) Including color cable 4(32CH) 1 36 1 Including on-board software, driver and manual USB Cable Service warranty card Package acceptance form 1 1 09 Operating manual 1.
LA-series Operating Manual 1-5 Optional Components ● ● 8CH signal capture box (including color cable) ● signal capture hook 8CH signal capture cable ● -5- USB A to B Cable
LA-series Operating Manual Chapter 2 Assembly 2-1 Check prior to Assembly 1. The ex-factory instruments are already tested and checked. Please unpack and check firstly to ensure the instruments are free of damage during the transportation. 2. Check if the accessories and quantities are complete according to the list of accessories, otherwise notify the company or local agent immediately for timely handling. 2-2 Operating Location 1.
LA-series Operating Manual 2-4 Utilization of Backup Fuses Please make sure of the fuse specification if backup fuse is required: ● Fuse: 250V/2Amp. ● Another kind of fuse may be added into the fuse holder. Note! the fuse shall be procured and replaced based on the specifications, otherwise the instrument may be damaged.
LA-series Operating Manual 2-5-2 Description of components ● Signal capture box(including color cable) -> Insert the connector into Port ● Signal detection cable ->Insert the connector into the signal capture box. Single-pin terminal is inserted into termianl or test point of DUT.-> * Note! don’t place the signal capture cable directly onto the port of LA in order to avoid damage of LA. * Send to the factory for repair or replacement in the case of breakup or damage.
LA-series Operating Manual 2-6 Startup Test 1. Insert one end of power cord into AC-in at left of LA, And the other end into AC power socket. 2. Press "1" of AC power switch and change to "1" from "0". ← AC power switch ←AC power socket 3. Press again orange power control switch at right upper side of LA front panel, the red LED is highlighted. Once upon buzzering one time, LA starts self-test, and enters into the following initial display, indicating the completion of startup test. 4.
LA-series Operating Manual Chapter 3 Description of Panels 3-1 Description of Front Panel -10-
LA-series Operating Manual 3-1 Description of Front Panel No.
LA-series Operating Manual 3-1-1 Menus(item) control keys are described below: 1. Channel setting key: After pressing the channel keys, 6 functional boxes at right-hand correspond to F1~F6 function keys. Adjust to preset channel using the shuttle (whitened location). Functionalities of functional keys: --Channel ON/Off: set ON or OFF of the channels --Bus: set the bus; the system is available with 16 BUS displays, each of which has 2 Channel~16 Channel.
LA-series Operating Manual Functional descriptions: 1. Bus: the system is available with 16 BUS, select the intended BUS with shuttle. 2. Source: source channel. select the intended channel with shuttle , then put the selected channel into Objective by key. 3. Objective: confirmed channel. If channel is provided, select the intended channel with shuttle ,then put the selected channel into Source by key.
LA-series Operating Manual - Move Down: move down CH selected in Objective, as shown below: - Back: back to the preceding option. - Disp Value On/Off(numerical display): display value on waveform, shown as ON or OFF. - Reorder 0←→31(reorder 0 ~31): reorder and display channels and names(CH0~CH31). - Name On/Off: ON/OFF of channel name, ON during startup, and OFF during switching-off. As shown above, the operating state is shown as red color. - Exit: exit from the setting state of channel.
LA-series Operating Manual 2. Change page key: a. Press this key and change to the second page when the number of channels or Buses on the screen exceeds 16, namely, switch back press again. b. The functional page can also be changed by "trigger". 3. Auto-scale key: Search the captured existing signal waveforms through auto-search by LA, which are automatically displayed on LCD about 3~5 seconds *Signals below 10Hz cannot be searched. 4.
LA-series Operating Manual 5. File setting key: Press this key, F1~F4 can be used to manage the file function items: - Load Setting: 5 functional keys (F1~F5) are available. The system setting data must be firstly stored and then accessed. - Save Setting: 5 functional keys (F1~F5) are available for the user to store the system setting data. - Load Waveform: 5 functional keys (F1~F5) are available to access the stored system settings and signal waveform data.
LA-series Operating Manual 6. Utility key: Press this key, F1,F2,F3,F4,F5 are available. - Display Mode: waveform display or character display mode are available. - B/W(background of LCD is black or white): setting ON/OFF of LCD background. - PC-Link: the captured signal waveform could be sent via USB2.0 to PC port for zoom-in, visualization and printing. For the operation in relation to on-line PC, refer to Chapter 5. - RAM Size(memory size): the memory size could be adjusted by turning shuttle.
LA-series Operating Manual 7. Main delays setting key: F1~F3: F1: 10%, F2: 50%, F3:90%; for F2: 50%, it indicates that, based on trigger center axle and "T" axle, the benchmark is displayed at 50% of LCD display, and the percentage could be adjusted by shuttle. 8. Grid ON/OFF key: Press this key, the vertical scale of time axle’s broke line on LCD could be controlled for ON/OFF. 9. logic levels key: Press this key, the logic levels could be changed in combination with F1~F3 function keys.
LA-series Operating Manual 3-1-2 Capture control keys are described below: 1. Single trigger key: Only a single signal is captured and displayed. "pat" label is displayed at right upper hand of LCD, signal capture is stopped once completion of capture. 2. Auto-store key: The test results are overlapped and displayed on the screen. In such case, flashing "pat" label is displayed at right upper hand of LCD. * Stop this function by pressing . 3.
LA-series Operating Manual 3-1-3 Trigger control keys are described below: 1. Internal/external clock key: Used to switch internal/external modes. - Internal mode, trigger with internal frequency signal of LA, namely, the so-called asynchronous trigger mode or time mode. - External mode, trigger with external frequency signal of LA, namely, the so-called synchronous trigger mode or state mode. * The character label of selected mode is blue.
LA-series Operating Manual 2. Trigger key: Press this key, F1~F5 represent the optional trigger modes: - a. Continuous(continuous ON/OFF): set multiple-PAT trigger conditions as continuous/discontinuous. (in combination with " "key). b. Trigger Counter: set the range of trigger between 1~255(in combination with " " key). - a. Pat 1: select and perform Pat 1 trigger condition (in combination with " " key). b. Pulse Width: select pulse width as a trigger condition, Set which channel ">", "=", "<".
LA-series Operating Manual 3-1-4 Character key: Numerical and character key: .。· 1. Numerical key: 0,1,2,3,4,5,6,7,8,9. Press these keys to display English characters and other symbols contained in the numerical keys. 2. : delete unnecessary characters. 3. : switch the cursor A~cursor F. 4. : control the cursor rightwards; " leftwards. 5. : enter key. Enter into delay and search functions by pressing this key under non-channel setting.
LA-series Operating Manual 3-1-5 Special function keys: Up/down, left/right orientation keys: 1. Press〝Λ〞,〝Ⅴ〞, you can change the zoom-in/out of waveform, and select logic levels, trigger PAT and Sample Rate. 2. Press〝<〞 , 〝>〞, you can visualize by adjusting the left and right position of waveform, or shift and select PAT under trigger. Shuttle key: the shuttle key has the following functions: 1. Select the channel to be displayed or to be preset. 2.
LA-series Operating Manual 3-1-6 Signal input port 1. Port1~4, four ports for signal capture input. Port 1: CH 0 ~ CH 7 Port 2: CH 8 ~ CH15 Port 3: CH16 ~ CH23 Port 4: CH24 ~ CH31 Every Port has a junction box and a group of 10 signal capture cables (including GNDx2) 2. Reference square-wave clock: The system is available with: a ~5V square-wave clock for calibration and self-measurement of channels.
LA-series Operating Manual 3-2 Description of Lateral Interface 1 2 3 4 1. USB2.0 interface, PC-Link. 2. AC input power switch (0: OFF, 1: ON): the power control switch of the front panel could be controlled efficiently only when AC power switch is firstly turned ON. 3. AC fuse holder. 4. Three-wire socket of AC power cord.
LA-series Operating Manual 3-3 Description of Rear Panel 1. Tips on precautions 2. Product names and serial number 3.
LA-series Operating Manual Chapter 4 Operating Instructions 4-1 Startup of System Power 1. Confirm if AC power complies with input power of the instrument. 2. Turn AC power switch at left-hand of LA from “0” to “1”, so AC power is turned on. 3. The red LED indicator is ON press power control switch at right upper hand of LA front panel. If buzzer gives a "buzz", "system test", etc, will be displayed on LCD. 4.
LA-series Operating Manual 4-2 Shortcut Operation LA-2050 shortcut flow graph LA-2050 shortcut operation description Check the instrument models, appearance and components Check if the instrument is operated properly. a.Check if the models and components are completely provided. b.Check if the instrument appearance is free of damage. Confirm if input power specification complies with instrument power SPEC Confirm the specifications of input power. a.
LA-series Operating Manual LA-2050 shortcut flow graph Set CH numbers and names ( 0~32 ) LA-2050 shortcut operation description 4. Set CH numbers and names: a. After pressing “Channel” key, you can set with F1~F5: channel numbers, ON/OFF and channel names. b. Adjust the cursor to the intended position with shuttle key. c. Set various parts where required. d. Exit from CH setting by pressing F6. Set the logic levels 5. Set the logic level: a.
LA-series Operating Manual LA-2050 shortcut flow graph LA-2050 shortcut operation description 6. Set trigger state of CHs: a.Enter into trigger state setting by pressing “trigger” key. Set trigger states of CHs b.Shift to CHs by shuttle key, the red position means the setting position. c. Set by selecting Low, High, X, , where necessary. , d. Exit from setting by pressing F6. Capture the trigger signal OK ! 7.Capture trigger signal: a.
LA-series Operating Manual 4-3 Setting of Channel Number and Name 1. Enter into the following channel setting state interface by pressing After selecting the channels to be modified with shuttle, F1~F6 are displayed, with the operations described below: Used to set or change the channel name. The following appears at left lower part of LCD: Name: BUS00 ↑ Note: 1. BUS00 under Name represents the name of CH0. 2."↑" indicates the character entry position of channel name.
LA-series Operating Manual 4-4 Adjustment of Logic Levels of Channels 1.Enter into logic levels setting display by pressing Press to select the Port(blue > indicates the existing Port), then adjust the voltage using shuttle key . ● F1,F2,F3 adjustable function keys: Press , it indicates "logic level" is TTL, and default: 1.5V, the voltage of signal to be captured could be adjusted using shuttle or "V", 〝Λ〞. Press , it indicates "logic level" is CMOS, and default: 2.
LA-series Operating Manual 4-5 Setting of Trigger State of Channels 1. Enter into trigger state displays of channels by pressing 2. The optional trigger states are displayed at right hand of LCD: F1: continuous On/Off F2 : PAT 1 F3 : PAT 2 F4 : PAT 3 F5 : OR PAT F6 : Exit 3. Change to the second page by pressing F1: trigger number (trigger after 1~255 times) F2: pulse width The trigger number may be adjusted by "shuttle" or "numerical key" and " → ","←".
LA-series Operating Manual Pulse width: the external signal pulse width could be set for trigger. Optional functions of pulse width: 1. CH: select trigger channel (S0~S31): select the measured channel with shuttle. 2. NAME: display channel name (the names of selected channels are displayed) 3. LEVEL: set the trigger logic level by selecting〝H〞or〝L〞with shuttle. 4. Selection of trigger condition: ">" (more than), "=" (equal), "<"(less) Trigger conditions could be adjusted using shuttle. 5.
LA-series Operating Manual The Trigger steps are descried below: 1.Enter into trigger state display of channels by pressing 2. After entering into PAT一(Pat 1) by pressing F2, the existing states of channels are displayed at lower part of LCD: Pat 1 xxxxxxxx xxxxxxxx xxxxxxxx xxxxxxxx * Initial state: 〝Don't Care〞. 3. Change the position of cursor to the channel with shuttle, "x" at lower part of LCD indicates the position of setting.
LA-series Operating Manual Press , the setting of channel is changed to " high logic level is turned to low logic level. ", namely, trigger when 4. Make the shuttle cursor point to CH1. Press , setting of CH1 is changed to H logic level, and so on. The settings are shown below: Pat 1 xxxxxxxx xxxxxxxx xxxxxxxx xxxxxxHL Exit from this setting by pressing Exit. Reference example of pulse width: If the frequency to be measured is 50KHz, the cycle width is 20uS, and pulse width 10uS.
LA-series Operating Manual The setting steps: 1. With in the main display, adjust the sample rate by pressing "﹀","︿". * When Sample Rate is adjusted, it requires faster than adjusting external pulse width. Adjust to 100ns 2.
LA-series Operating Manual 3. Change to the second page by pressing 4. Select pulse width by pressing 5. Select channels (from S0~S31) using shuttle. 6. Set logic level as〝H〞(starting when external signal is High) 7. Set〝=〞(equal) 8. Adjust to〝010.000uS〞 9. Perform key to capture the waveform and display on LCD.
LA-series Operating Manual 4-6 Capture of Trigger Signal 1. After completion of settings, the capture hook of signal junction box is linked to the pin of objects. 2. Start to capture the triggered signal waveform after the object is energized. 3. 3 capture modes for trigger signal waveform are available where necessary. 4. Mode 1: press , it indicates it’s intended to capture a single trigger signal. After capture, the operation is stopped, and the waveform is displayed on LCD.
LA-series Operating Manual 4-7 Visualization of Captured Signal Waveform After capture of trigger signals by the preceding 3 modes, the signal waveforms of various channels could be visualized, where necessary. Change the page by pressing 1. There are 32 CHS that can be used and displayed by LA-2050, of which 16CH are displayed on LCD, and Page1 or Page2 is used to change the page and visualize the display. 2.Page 1:CH0~CH15 are defaults, Page 2:CH16~CH31. 3.
LA-series Operating Manual 4-8 Save and Load of Waveform Data Save waveform data 1. Enter into save waveform and load waveform by pressing "file setting" key, as shown below: F1: load setting F4: save waveform F2: save setting F6: Exit F3: load waveform, 2. Enter into save waveform by pressing F4 , wherein there are 5 groups of waveforms for saving; press F1 , it indicates waveform data is saved at the designated position of F1. 3. Back by pressing F6 4.
LA-series Operating Manual Save system setting As shown above: 1. Enter into save setting by pressing system setting data for saving; press the designated position of F1. 2. Back by pressing F1 , wherein there are 5 groups of , it indicates the data is saved at F6 3. Enter into load setting by pressing F1 , wherein there are 5 groups of system setting data for loading; press F1 , it indicates the system setting data is loaded at the designated position of F1. 4.
LA-series Operating Manual 4-9 Turn off Power of LA System 1. If it’s intended to switch off LA in the operating state, press off the power. to turn 2. The following appears at lower part of LCD: Save Boot Wait . . . . It’s closed 3~4 seconds after relevant system data is saved. 3. Turn off power of AC or remove AC power cord: If the instrument is in idle state for a long time, turn off AC power switch at left hand of LA from "1" to "0".
LA-series Operating Manual Chapter 5 Operation of PC-LINK 5-1 Installation of PC-LINK 1. PC- Link is mainly used to send the captured signal waveform of LA to PC for saving, visualization, search and printing. 2. PC- Link has a communication interface USB 2.0, so USB driver shall be installed. 3.
LA-series Operating Manual 5-2 Operation of PC-LINK 1. If trigger signal waveform is already captured at LA-2050 terminal, this signal waveform could be loaded to PC via USB 2.0. 2. Press at LA terminal, the following appear at right hand of LCD: F1~F5 function options, select F3 as PC Link function, and press F3, indicating this host terminal is ready to send waveform data. 3.
LA-series Operating Manual 5-3 Software System Requirements CPU: 3.0GHz or higher. Ram: at least 512MB. HD: 50MB. Port supporting USB2.0. 5-4 Software Installation Step1: close all running programs. Step2: place CD in the optical drive, and install by the following steps.
LA-series Operating Manual If Optical disc isn’t automatically performed, press "Start" button of Windows, then press "Execute". Enter "D﹕\setup.exe" (assuming optical drive is: "D﹕\") in "Activate" field. Step3: press "Next", the option display of installation path will appear: Step4: start installation by pressing “Next”, and finally press "Finish".
LA-series Operating Manual 5-5 Hardware Installation Step1: Logic Analyzer is linked to PC via USB; in the case of first installation, the following display will appear: Select “yes, only now”. Step2: the following display appears press "Next".
LA-series Operating Manual Step3: the following display appears press "Next" Select “this location is included during search”, and press "review".
LA-series Operating Manual There is a “Driver” folder under the installed data folder (preset as: c:\program files\Logic Analyzer V2.0\); select the data and press "Confirm". Step4: after pressing "Confirm", select "Next" to start installation of the intended Driver. The following warning window will appear during installation: Select “Continue to install” for completion of Driver installation.
LA-series Operating Manual Step5: after completion of installation, press “Finish”.
LA-series Operating Manual 5-6 Software Execution Method 1: 1. Press "Start" function key, select "all programs". 2. Start the software by selecting LogicAnalyzer V2.0 -> LogicAnalyzer V2.0. Method 2: start the software by clicking directly the software tag on the desktop.
LA-series Operating Manual 5-7 Interface 5-7-1 Model Selection Select the types by pull-down menu. 5-7-2 Operating Interface (1) Window 1. Waveform window A: Functional option list. B: Tool list. C: Message list. D: Display of channel names. E: Display of trigger mode. F: Waveform display area.
LA-series Operating Manual 2. State Mode A: State mode display area. B: Display of channel names. C: Display of trigger mode.
LA-series Operating Manual (2) Menu 1. File menu Load: load the files saved in PC. Save: save the existing data into file. Import: import the data of host computer to PC. Export: export PC data to the host computer. Capture: capture the existing display into file. Report: put the existing data into report. Print: print the existing waveform. Exit: close the existing programs. 2. View menu Zoom In: zoom-in waveform. Zoom Out: zoom-out waveform. Grid Style: grid style switching.
LA-series Operating Manual 4. Tool menu a.TriggerEdit: set trigger conditions. b.Channel/Bus Edit: Set Channel/Bus. c.GOTO Cursor: Position the cursor. 5.Help menu Default: ex-factory setting. About: software information.
LA-series Operating Manual (3) List of Operations 1. File function list A: load. B:save. C: capture display. D: print. 2. Advanced function list A: call BUS setting dialog box. C: single sampling. E: stop. B: call Trigger setting dialog box. D: continuous sampling. F: auto-search. 3.Utility operation list A: zoom-out. C: new cursor. E: grid style. B: zoom-in. D: cursor spacing display. F: background white.
LA-series Operating Manual 5-8 Detailed Operations 5-8-1 File Functions (1) Save Method 1: Select "File" menu and then "Save". Method 2: Click the disc pattern on the tool list. (2) Load Method 1: Select "File" menu and then "Load". Method 2: Click the file folder pattern on the tool list. (3) Export Connect Logic Analyzer with PC via USB. Then, select "File" menu and Export, thus exporting data to Logic Analyzer. (4) Import Connect Logic Analyzer with PC via USB.
LA-series Operating Manual (5) Report Print-out Select "File" menu and then "Report Out" to call "Report Out Dialog Box". Report Out Dialog Box: A: Route selection B: File selection C: File name entry box D: Name of test personnel E: Selection of export range (6) Capture Method 1: Select "File" menu and then "Capture" to call capture dialog box. Method 2: Click and select the camera pattern on the tool list to call the capture dialog box.
LA-series Operating Manual (7) Print Method 1: Select "File" menu and then "Print" to call print dialog box. Method 2: Click and select the printer pattern on the tool list to call the print dialog box.
LA-series Operating Manual 5-8-2 Waveform capture (1) Auto-search Method 1: Select "Run/Stop" menu and then "Auto Scale" to search automatically waveform data and set the system parameters. Method 2: Click the display pattern on the tool list to search automatically waveform data and set the system parameters. (2) Single sampling Method 1: Select "Run/Stop" menu and then "Single Run" to capture individually the waveform data.
LA-series Operating Manual 5-8-3 Waveform analysis (1) Grid style Method 1: Click "View" menu and then "Grid Style" to change the grid display mode. Method 2: Click the grid pattern on the tool list to change the grid display mode. (2) Background color Method 1: Click "View" menu and then "B/W" to change the background color. Method 2: Click the B/W pattern on the tool list to change the background color. (3) Waveform zoom-in Method 1: Click "View" menu and then "Zoom In" to zoom-in the waveform.
LA-series Operating Manual (5) Waveform shift Drag the scroll axle below the waveform display to shift the waveform leftwards or rightwards. Adjust the position of waveform Press and drag the left mouse button on the left side to adjust the position of waveform. The selected channel is highlighted by a red box.
LA-series Operating Manual 5-8-4 Settings 1. Setting of sampling frequency Click the pull-down menu of the tool list under "waveform window". 2. Setting of memory depth Click the pull-down menu of the tool list under "waveform window". *In the case of bigger memory depth and lower sampling frequency, it takes longer time to capture data. In such case, “Auto” option could be selected to enable automatic setting of memory depth for faster data capture.
LA-series Operating Manual 3. Setting of Channel/ BUS Method 1: Click "Tool" menu and then "Channel/Bus Edit" to call Channel/BUS edit dialog box. Method 2: Click right mouse button in waveform display under waveform mode, select "Advance BUS" from the pull-down menu to call Channel/BUS edit dialog box. Click right mouse button in data area under state mode, and select “ Advance BUS” from the pull-down menu to call Channel/BUS edit dialog box.
LA-series Operating Manual (1) BUS setting display A: Channel state; “Mode”, “name” and “bits in channel” from left to right. B: Channel/BUS setting zone C: PORT voltage display D: Function key (2) Channel Setting A: Current state; This setting is valid only when the channel is BUS. Bits indicate the quantity of Bit(Channel) in the channel, 1 is single channel, 0 is shut-off; the others bigger than 1 is BUS.
LA-series Operating Manual (3) Bus Setting BUS or Channel is set by dragging with left mouse button at Channel/BUS setting zone. A: Current state; Auto represents auto-judgement mode. User represents user-defined mode. Bits indicate the quantity of Bit(Channel) in the channel, 1 is single channel, 0 is shut-off; the others bigger than 1 is BUS. B: BUS display under Auto mode; C: BUS display under User mode; 1.
LA-series Operating Manual (4) New channel A: Press “Add” at the lower position. B: Add a new void channel at the lowest position of all channels for the user.
LA-series Operating Manual (5) Delete channel Firstly, select the channel to be deleted. A: Click “Delete” at lower position. B: The selected channel will be deleted.
LA-series Operating Manual (6) Erase channel Firstly, select the channel to be erased. A: Click “Erase” at lower position. B: The selected channel will be erased.
LA-series Operating Manual 4. Trigger setting (1) Simple Trigger setting Method 1: Call the rapid setting list by double-clicking "Trigger/Pattern" field of the channel in the "waveform window". Method 2: Call the rapid setting list by double-clicking "Sample/Trigger" field of the channel under "state mode".
LA-series Operating Manual (2) Advanced Trigger setting Method 1: Click "Tool" menu and then "Trigger Edit" to call Channel/BUS Edit dialog box. Method 2: Press right mouse button of waveform display under waveform mode, select "Advance Trigger" from pull-down menu to call Trigger Edit dialog box. Press right mouse button at data zone under state mode, select “Advance Trigger” from pull-down menu to call Trigger Edit dialog box.
LA-series Operating Manual (I) Trigger setting page A: Pattern mode switching. B: Trigger Counter and continuous/discontinuous setting. C: Erase key. (II) Erase Pattern setting Press the lower “Erase” key to erase all existing Pattern settings; All Trig states reset to “Don’t Care" under Pattern mode. The settings are set as "CH0" "High" "=" "0" "ns" under Pulse width mode.
LA-series Operating Manual (3) Pattern (I) Pattern switching Click the left Pattern key for mode switching. (II) Setting of Pattern A: Pattern state zone. To change the trigger mode of Pattern, call the state menu in the state zone by right mouse button: The state of channel can be set by clicking the required state.
LA-series Operating Manual 1. In the case of “Low”, “High” and “Don’t Care", press and hold the left mouse button on the state, then drag leftwards and rightwards to set rapidly the trigger state. 2. In the case of “Raising” and “Falling”, press and hold the left mouse button on the state, then drag leftwards and rightwards to shift the position of “Raising"and "Falling". * Note: only “Low”, “High” and “Don’t Care" trigger state can be selected under Pattern2 and Pattern3.
LA-series Operating Manual (4) Pulse width A: Switch to PulseWidth setting page by clicking “Pulse Width” button of left Pattern column. B: Select the channel. (CH0~CH31) C: Channel name(auto-display). D: Select logic level. (Low, High) E: Select condition.
LA-series Operating Manual (5) Continuous and discontinuous A: Click “On” and “Off ” in “Continuous” at right information column. * This function is effective only in the case of “ Pattern2” and “Pattern3”. * If continuous/discontinuous is set as "On", trigger is effected only Pattern2 and Pattern3 match continuously Pattern state, without mixture of other states; otherwise no continuous data is required for triggering.
LA-series Operating Manual 5. Voltage setting Call the voltage setting dialog box by clicking the upper voltage display zone with left mouse button in BUS setting page. A: Click this option, all Port voltages are set consistently, otherwise set individually. B: Standard voltage setting; “TTL”, “ECL” and “CMOS”. C: User define; user-defined voltage logic level.
LA-series Operating Manual 6. Delay Call delay entry dialog box by clicking “Delay” key on the tool list. A: Select unit. BKSP: delete a character by cursor. Clear: clear the contents in the entry box. "←" , "→": left and right shift of cursor.
LA-series Operating Manual 7. Change the channel name Method 1: Under the “waveform window”, double click the channel name to call the name setting dialog box. Method 2: Under the "state mode", double-click the channel name to call the name setting dialog box.
LA-series Operating Manual Method 3: Under “BUS” setting dialog box, double-click the channel name to call the name setting dialog box. Name setting dialog box: BKSP: delete a character by cursor. Clear: clear the contents in the entry box. Caps: big/small number switching. Shift: character/symbol switching. "←" , "→" : left and right shift of cursor.
LA-series Operating Manual 8. Switch BUS numerical display mode Method 1: Under the "waveform window", double click the "Sample/Trigger" field of the channel to call the numerical display options. Hexadecimal: display in hexadecimal system. Decimal: display in decimal system. Binary: display in binary system. Method 2: Under the "state mode", double click the "Sample/Trigger" field of the channel to call the numerical display options. Hexadecimal: display in hexadecimal system.
LA-series Operating Manual 9. Cursor setting (1) Add new cursor A new cursor is added into "waveform window" or "state mode" by clicking "Add" icon on the tool list. * The quantity of cursors is limited to 26. (2) Delete cursor Call the function options by clicking the right mouse button on the cursor to be deleted. Delete the selected cursor by clicking “Delete”. (3) Shift cursor Press and hold left mouse button on the intended cursor, drag leftwards and rightwards to shift the position of cursor.
LA-series Operating Manual (4) Search the cursor Method 1: call the function options by clicking right mouse button on "waveform display zone" in "waveform window". Method 2: Call the function options by clicking right mouse button on "state display zone" in "state mode".
LA-series Operating Manual Method 3: Click "Tool" menu and then "GOTO Cursor" to call Channel/BUS edit dialog box. Click “GOTO Cursor” to call the cursor and search the dialog box. Select the cursor to be searched from the pull-down menu. Enable the central point to align the trigger cursor by clicking “GOTO Trigger”. Enable the central point to align the starting point of data by clicking “ GOTO Begin”. Enable the central point to align the ending point of data by clicking "GOTO End".
LA-series Operating Manual 5-9 Shortcut Flow Process 1. Link the Logic Analyzer to computer. 2. Switch the Logic Analyzer to PC-Link mode. 3. Perform Logic Analyzer PC software. 4. Select device. 5-1 Select auto-search key on the tool list. Auto-search key on operation column “Auto-search” can detect automatically if it’s possible to capture signals, Auto-scale sampling frequency and memory depth.
LA-series Operating Manual 6-1. Call the name setting dialog box by double click the name display zone. 6-2. All signals are named using dialog box.
LA-series Operating Manual 7-1. Call simple trigger options by double click “Trigger / Pattern” field. 7-2. After completion of setting, the waveform is triggered according to the set trigger state.
LA-series Operating Manual 8-1. If auto-search is not used, the channel is set by pressing right button in the waveform display zone. 8-2. BUS combination and Channel assignment are performed by means of dragging.
LA-series Operating Manual 8-3. Call the voltage setting dialog box by clicking left mouse button on the voltage display zone. 8-4. Set voltage in the dialog box. 8-5. Switch to the trigger page by clicking the upper Trigger subpage.
LA-series Operating Manual 8-6. Set advance trigger in the Trigger page. 8-7. Return to the waveform window, and capture the waveform by pressing dual-arrow on the upper tool list.
LA-series Operating Manual 8-8. After data capture, set trigger by repeating the above 6~7 steps. 9. Return to waveform window, set the sampling frequency by clicking “DIV ” pull-down menu on the tool list.
LA-series Operating Manual 10. Set the memory depth by clicking “Memory Depth” pull-down menu on the tool list. 11. Enter Delay value by click “Delay” key on the tool list. Enter Delay value into the dialog box with Delay.
LA-series Operating Manual Chapter 6 Calibration Description It’s advisable to calibrate the instrument since an error between setting and import/export value exists due to certain factors after a period of time (often 1 year). 6-1 Calibration Mode Two calibration modes are described below: 1. The user shall apply for calibration and then send the instruments back to the factory; this company provides ex-factory calibration report. 2.
LA-series Operating Manual Chapter 7 Maintenance, Repair and Utilization Description Maintenance and repair is divided into: maintenance by users and maintenance by factory. 7-1 Maintenance by Users 1. The users may maintain the devices without removing the instrument housing. 2. If, in the case of correct AC input power, the instrument is still disabled when PowerSwitch at left panel is “1” and Power key on the frontal panel is pressed, the fuse is already burnt.
LA-series Operating Manual 7-2 Maintenance by Factory The components shall be maintained or replaced by the manufacturer or distributor in either of the following cases: 1. After the fuse in good working order is replaced and AC power is confirmed, the instrument cannot still be opened normally. 2. The system is still maintained at open interface after startup, but it’s unable to log in operable interface. 3. Continuous operation cannot proceed in the presence of error message. 4.
LA-series Operating Manual 7-3 Troubleshooting 1. Description: please read carefully the operating instructions if certain problems or doubts are encountered when logic analyzer is employed. 2. Operating questions and countermeasures: Q1: In the event of Power on during startup, why the power indicator isn’ t highlighted, and it’s unable to display normally? Ans: Check if power cord is fastened securely, the fuse in fuse holder is normal, and AC power switch is at "1".
LA-series Operating Manual Q6: How does signal capture line and loop of every channel work normally? Ans: Connect the signal capture line of every channel to the reference square-wave of LA, and observe if the square-wave can be captured. Q7: During normal operation of circuit, can the waveform be captured after triggering? (are solid materials connected normally) Ans: Confirm if the set LA trigger state, mode and logic level conform to the specification characteristics of test circuit.
LA-Series Operating Manual Appendix A: I2C series bus decode 1. Start I2C Decode function Method 1 Select “Tool”→“Decoder”→“I2C” to start the I2C decode function. Method 2 Click and select the I2C pattern on the tool list to start the I2C decode function.
LA-Series Operating Manual 2.
LA-Series Operating Manual 3.
LA-Series Operating Manual 4. Display field Field Packet Status Address Msg. Time Idle Time Dir Length Data Function Packet counter S: Start P: Stop Display of address Message continuous time Message idle time Write or load Packet length Data 5. Setting Procedures Step 1: Setting channels of SCL and SDA Step 2: Setting of text mode.
LA-Series Operating Manual Step 3: Setting of address bit and data bit. Step 4: Setting of READ BIT level high or low. Step 5: Setting of ACK BIT level high or low. Step 6: Select “Start” button to analyze I2C message.
LA-Series Operating Manual Appendix B: SPI series bus decode 1. Start SPI Decode function Method 1 Select “Tool”→“Decoder”→“SPI” to start the SPI decode function. Method 2 Click and select the SPI pattern on the tool list to start the SPI decode function.
LA-Series Operating Manual 2.
LA-Series Operating Manual 3.
LA-Series Operating Manual 4. Display field Field Packet Error bit Length Data Msg.Time Abs. Time Idle Time Function Packet counter Error bit Package length Data Message continuous time Continuous time plus idle time Message idle time 5. Statistics Field Field Min MSG. Time Max MSG. Time Avg MSG.
LA-Series Operating Manual 6. Setting Procedures Step 1: Channel setting: Setting the decode channel and the SS level high or low. Step 2: Enable MISO or/and MOSI. SStep 3: Setting channels and parameters of MISO and MOSI.
LA-Series Operating Manual Step 4: Setting of text mode. Step 5: Select “Start” button to analyze SPI message.
LA-Series Operating Manual Appendix C: UART decode 1. Enable Decoder a: Select Tool → Decoder → UART to enable UART Decoder b:Click UART icon to enable UART Decoder.
LA-Series Operating Manual 2. Interface A: Parameter setting B: Waveform and data.
LA-Series Operating Manual 3.
LA-Series Operating Manual 4. Display Block Block Packet Status Data Parity Function Packet counter S: Start P: Stop Data Result of Parity check 5.
LA-Series Operating Manual Step 3: Set data Bits Step 4 Set Parity But Step 5: Set Stop bits Step 6: Set LSB or MSB -114-
LA-Series Operating Manual Step 7: Set Logic Level Step 8: Set text mode Step 9: Press start to decode -115
LA-Series Operating Manual Appendix D: CAN decode 1. Enable CAN Decoder a: Select「Tool」→「Decoder」→「CAN」to enable CAN Decoder b: Click CAN icon to enable CAN Decoder 2.
LA-Series Operating Manual 3.
LA-Series Operating Manual 4. Display Block Block Basic ID Extended ID RTR IDE Ctr Field Data CRC ACK Function Show Basic ID Show Extended ID Show RTR value Show IDE value Show Control Field value Show DATA content Show CRC value Show ACK 5.
LA-Series Operating Manual Step 3: Set hide or display extension ID Step 4: Press start to decode 5.