106 WA1K2K-OM-E RevB
(3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz -
2.7 GHz)
EN 61000-4-4:2004** Electrical Fast Transient/Burst.
(1 kV on power supply lines, 0.5 kV on I/O signal data and control
lines)
EN 61000-4-5:2006** Surges.
(1 kV AC Mains, L-N, L-PE, N-PE)
EN 61000-4-6:2007** RF Conducted Electromagnetic Field.
(3 Vrms, 0.15 MHz - 80 MHz)
EN 61000-4-11:2004† Mains Dips and Interruptions.
(1 cycle voltage dip, 100% short interruption)
* In order to conform to Radiated Emissions standards, use properly
shielded cables on all I/O terminals.
** Meets Performance Criteria "B" limits during the disturbance; product
undergoes a temporary degradation or loss of function of performance
which is self recoverable.
† Meets Performance Criteria "C" limits during the disturbance; product
undergoes a temporary degradation or loss of function of performance
which requires operator intervention or system reset.
WARNING
This is a Class A product. In a domestic environment this product may
cause radio interference, in which case the user may be required to take
appropriate measures.
Low-Voltage Directive
EN 61010-1:2001
Safety requirements for electrical equipment for measurement, control,
and laboratory use.