User manual

Table Of Contents
Operator's Manual
6 WA1K2K-OM-E RevB
AC Power Source
100 to 240 VAC (±10%) at 50/60/400 Hz (+10%); Installation Category:
300V CAT II
No manual voltage selection is required because the instrument
automatically adapts to line voltage.
Power Consumption
< 50 Watts for all models (depending on accessories installed - probes,
USB devices, etc.).
Power consumption in Standby Mode: < 5 Watts.
Power and Ground Connections
The instrument is provided with a 10A/250V 18AWG rated grounded
cord set containing a molded three-terminal polarized plug and a
standard IEC-60320 (Type C13) connector for making line voltage and
safety ground connections. The AC inlet ground terminal is connected
directly to the frame of the instrument. For adequate protection against
electrical shock hazard, the power cord plug must be inserted into a
mating AC outlet containing a safety ground contact.
WARNING - Electrical Shock Hazard
Only use the power cord provided with your instrument.
Any interruption of the protective conductor inside or outside of the
oscilloscope, or disconnection of the safety ground terminal creates a
hazardous situation. Intentional interruption is prohibited.
In Standby mode, the oscilloscope is still connected to the AC supply. The
instrument can only be placed in a complete Power Off state by flipping
the main power switch on the rear of the instrument (if provided) to the
off (Zero) position or by removing the power cord. In cases where no
main power switch is provided, removing the main power cord is the
disconnect device.